Device recall enforcement reports

Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/130; Material Number: M001195440

Boston Scientific Cork, Ltd. · The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcathe…

Class IIOngoingZ-0243-2027
Status
Ongoing
City
Cork
Country
Ireland
Classification
Class II
Product type
Devices
Event ID
99906
Recalling firm
Boston Scientific Cork, Ltd.
Address 1
Boston Scientific Limited, Business And Technology Park, Model Farm Road
Postal code
N/A
Voluntary mandated
Voluntary: Firm initiated
Initial firm notification
Letter
Recall number
Z-0243-2027
Product description
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/130; Material Number: M001195440
Product quantity
620
Recall initiation date
20260916
Center classification date
20261022
Report date
20261028
Distribution pattern

Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.

Reason for recall

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

Code info

Material Number: M001195440; UDI-DI: 08714729839514; Batch Numbers: 38181961, 38198832, 38224582, 38259483, 38324070, 38460469, 38560455, 38560456, 38595654, 38649857, 38745194, 38745195, 38810586, 38810591, 38904805, 39092267, 39092272, 39143222, 40470259, 40478976, 40478977, 40492318, 40497671, 40497858, 40502038, 40502039;