Device recalls

MiSeq Dx Instrument REF: DX-410-1001 / 15036706, FOR IN VITRO DIAGNOSTIC USE

Illumina, Inc. · (Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, serv…

Open, ClassifiedLabeling designZ-0321-2027
Open FDA's page for this record
Cfres ID
222975
Product res number
Z-0321-2027
Event date initiated
2024-05-08
Event date posted
2026-10-30
Recall status
Open, Classified
Res event number
99957
Product code
PFF
Product description
MiSeq Dx Instrument REF: DX-410-1001 / 15036706, FOR IN VITRO DIAGNOSTIC USE
Code info
All Lots/UDI: 00816270020002
Firm FEI number
3003218906
Recalling firm
Illumina, Inc.
Address 1
5200 Illumina Way
City
San Diego
State
CA
Postal code
92122-4616
Additional info contact
Christine Douglass 858-202-4500
Root cause description
Labeling design
Product quantity
1193 units
Reason for recall

(Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission.

Action

On May 8, 2024, Illumina, Inc. issued a Urgent Medical Device Recall via email. Illumina ask consignees to take the following 1. Download & read this Cybersecurity Guidance document. The full hyperlink is below. https://support.illumina.com/support-content/cyberguidance.html 2. For Each Product Connected to the Customer Network: work with your internal IT department to determine which secure protocol works best in your environment and to implement the appropriate configuration change. Note: a change to the network protocol affects both Dx and RUO modes simultaneously therefore there is no need to boot into each mode separately to take the necessary action. 3. Complete and return the Verification Form after carrying out all the steps in the instructions provided on your specific product(s) identified as affected in Table 1.

Distribution pattern

Worldwide - US Nationwide distribution including in the states and territory of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of AE, AR, AT, AU, BE, BG, CA, CH, CL, CN, CY, CZ, DE, DK, DZ, EG, ES, FI, FR, GB, Ghana, GR, GT, HK, HU, IE, IL, IS, IT, JP, KR, KW, LT, LU, LV, MA, MT, NG, NL, NO, NZ, PE, PH, PL, PT, QA, RO, RU, SA, SE, SG, SI, TH, TR, UA, VN, ZA.

Harmonized identifiers (openFDA)
Device name
High Throughput Dna Sequence Analyzer
Medical specialty description
Immunology
Regulation number
862.2265
Device class
2
K number
K170299, DEN130011
Registration number
3009875001, 3007287152, 3011104381, 3006296346, 3003335080, 3007102730, 3015529000, 3014150341, 3003673482, 3003218906, 3009207631
FEI number
3009875001, 3007287152, 3005095756, 3011104381, 3003335080, 3007102730, 3015529000, 3014150341, 3003673482, 3003218906, 3009207631
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.