VeriSeq Onsite Server V2 REF: 20047000
Illumina, Inc. · (Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, serv…
Open FDA's page for this record- Cfres ID
- 222979
- Product res number
- Z-0324-2027
- Event date initiated
- 2024-05-08
- Event date posted
- 2026-10-30
- Recall status
- Open, Classified
- Res event number
- 99957
- Product code
- PFF
- Product description
- VeriSeq Onsite Server V2 REF: 20047000
- Code info
- All Lots/UDI: N/A
- Firm FEI number
- 3003218906
- Recalling firm
- Illumina, Inc.
- Address 1
- 5200 Illumina Way
- City
- San Diego
- State
- CA
- Postal code
- 92122-4616
- Additional info contact
- Christine Douglass 858-202-4500
- Root cause description
- Labeling design
- Product quantity
- 201 units
- Reason for recall
(Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission.
- Action
On May 8, 2024, Illumina, Inc. issued a Urgent Medical Device Recall via email. Illumina ask consignees to take the following 1. Download & read this Cybersecurity Guidance document. The full hyperlink is below. https://support.illumina.com/support-content/cyberguidance.html 2. For Each Product Connected to the Customer Network: work with your internal IT department to determine which secure protocol works best in your environment and to implement the appropriate configuration change. Note: a change to the network protocol affects both Dx and RUO modes simultaneously therefore there is no need to boot into each mode separately to take the necessary action. 3. Complete and return the Verification Form after carrying out all the steps in the instructions provided on your specific product(s) identified as affected in Table 1.
- Distribution pattern
Worldwide - US Nationwide distribution including in the states and territory of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of AE, AR, AT, AU, BE, BG, CA, CH, CL, CN, CY, CZ, DE, DK, DZ, EG, ES, FI, FR, GB, Ghana, GR, GT, HK, HU, IE, IL, IS, IT, JP, KR, KW, LT, LU, LV, MA, MT, NG, NL, NO, NZ, PE, PH, PL, PT, QA, RO, RU, SA, SE, SG, SI, TH, TR, UA, VN, ZA.
- Harmonized identifiers (openFDA)
- Device name
- High Throughput Dna Sequence Analyzer
- Medical specialty description
- Immunology
- Regulation number
- 862.2265
- Device class
- 2
- K number
- K170299, DEN130011
- Registration number
- 3009875001, 3007287152, 3011104381, 3006296346, 3003335080, 3007102730, 3015529000, 3014150341, 3003673482, 3003218906, 3009207631
- FEI number
- 3009875001, 3007287152, 3005095756, 3011104381, 3003335080, 3007102730, 3015529000, 3014150341, 3003673482, 3003218906, 3009207631