Senographe Pristina 3D

GE Medical Systems, LLC · GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisitio…

Open, ClassifiedRadiation Control for Health and Safety ActZ-0328-2027
Open FDA's page for this record
Cfres ID
223534
Product res number
Z-0328-2027
Event date initiated
2026-08-28
Event date posted
2026-11-04
Recall status
Open, Classified
Res event number
100070
Product code
OTE
Product description
Senographe Pristina 3D
Code info
N/A
Firm FEI number
2126677
Recalling firm
GE Medical Systems, LLC
Address 1
3000 N Grandview Blvd
City
Waukesha
State
WI
Postal code
53188-1615
Additional info contact
262-544-3011
Root cause description
Radiation Control for Health and Safety Act
Distribution pattern
USA and OUS.
Reason for recall

GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).

Action

GE HealthCare will send a customer letter dated August 28, 2026, that provides safety instructions for immediate mitigation of the incorrect configuration of the collimator anode track / filter in manual mode during 3D+2D acquisitions. GE HealthCare will correct all affected Senographe Pristina VIA systems by upgrading the software at no cost to the customer. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

PMA numbers
P130020
Harmonized identifiers (openFDA)
Device name
Digital Breast Tomosynthesis
Medical specialty description
Unknown
Device class
3
PMA number
P160031, P130020, P080003, P140011
Registration number
3005810333, 9611343, 3008833258, 3003775006, 3023003446, 3019373577, 3012939903, 3040678329, 3031503890, 3013875850, 3009196578, 3029521953, 3012127758, 1423253, 3005877899, 3008496839, 3004977335, 3001722928, 3015276088, 3005496266, 3010703925, 3036666444, 3027339477
FEI number
3005810333, 3008833258, 3003775006, 3023003446, 1000514697, 3019373577, 3012939903, 3040678329, 3031503890, 3013875850, 3009196578, 3029521953, 1000284313, 3012127758, 3005877899, 3008496839, 3004977335, 3001722928, 3015276088, 3005496266, 3010703925, 3036666444, 3027339477