Device recalls

Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (…

SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION · The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and saf…

Open, ClassifiedProcess controlZ-0347-2027
Open FDA's page for this record
Cfres ID
223189
Product res number
Z-0347-2027
Event date initiated
2025-09-30
Event date posted
2026-11-03
Recall status
Open, Classified
Res event number
100044
Product code
OWB
Product description
Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),
Firm FEI number
3002808139
Recalling firm
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
Address 1
Nishinokyo Kuwabara -Cho 1 nishinokyo kuwabara cho nakagyo-ku
City
Kyoto
Country
Japan
Additional info contact
Mr. Joseph Zegarelli 8175-823-1305
Root cause description
Process control
Product quantity
118 unts.
Distribution pattern
US Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY.
Code info

Serial number/Shimadzu Medical Systems #/UDI: M1F0D3BF4001/ SMS20250167/ (01)04540217067145(11)250; M1F0CC4EB001/ SMS20250098/ (01)04540217067022(11)241; M1F0D27EA001/ SMS20250001/ (01)04540217067138(11)241; 41F0F43E7001/ SMS20250034/ (01)04540217068951(11)240; M1F0D3BE7001/ SMS20250248/ (01)04540217067145(11)240; M1F0D3BE6001/ SMS20240196/ (01)04540217067145(11)240; 41F0F40E5001/ N/A/ (01)04540217068937(11)240; 41F0F41E5001/ N/A/ (01)04540217068944(11)240; 41F0F43E5001/ SMS20240236/ (01)04540217068951(11)240; M1F0CD7E4001/ SMS20240162/ (01)04540217067046(11)240; 41F0F41E1001/ SMS20240062/ (01)04540217068944(11)240; 41F0F43DB002/ SMS20240101/ (01)04540217068951(11)231; 41F0F43DB001/ SMS20240100/ (01)04540217068951(11)231; M1F0CD7D4001/ SMS20230178/ (01)04540217067046(11)230

Reason for recall

The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.

Action

On September 30, 2026 Shimadzu Corp. issued a Urgent Medical Device Recall Notification to affected consignees via email. Shimadzu asked consignees to take the following actions; 1. If the bolts become loose, an abnormal noise will occur when moving the C-arm. Please check for abnormal noises during pre-study inspections. 2. If an abnormal noise is heard, please stop using the product and contact our service department. 3. Please complete the Medical Device Recall Customer Acknowledgement.

K numbers
K221922, K203535
Harmonized identifiers (openFDA)
Device name
Interventional Fluoroscopic X-Ray System
Medical specialty description
Radiology
Regulation number
892.1650
Device class
2
K number
K243735, K170130, K253670, K221344, K220342, K173631, K180473, K243884, K211191, K200917, K050489, K211780, K212442, K231700, K101868, K181550, K173314, K251523, K163715, K123508, K181804, K240465, K032046, K231669, K130638, K172646, K062647, K203281, K052202, K181834, K002198, K212383, K033244, K230787, K160147, K961374, K123529, K181415, K222625, K233230 +285 more
Registration number
3021059265, 3016438694, 3014391847, 2027467, 3038614464, 3010398867, 9681146, 3004785967, 3015223550, 3009196578, 3008632256, 3003202425, 3012761544, 1723170, 3001722928, 2243057, 3003768251, 3021561977, 3010703925, 3036666444, 3021027284, 3013184485, 3022378058, 3008284470, 3011713200, 3027373189, 3036255956, 3018094310, 8030233, 3006544299, 3016248538, 1225677, 3009973505, 1037955, 2133772, 3008363989, 3025394757, 2939520, 3030412764, 3014343100 +86 more
FEI number
3021059265, 3016438694, 1000114111, 3014391847, 2027467, 3038614464, 3002808139, 3010398867, 3004785967, 3015223550, 3009196578, 3008632256, 3003202425, 3012761544, 3001722928, 2243057, 3002980729, 3003768251, 3021561977, 3010703925, 3036666444, 3021027284, 3013184485, 3022378058, 3008284470, 3011713200, 3027373189, 3036255956, 3018094310, 3006544299, 3016248538, 3009973505, 1037955, 1000132370, 3008363989, 3025394757, 2939520, 3030412764, 3014343100, 1000188474 +86 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.