Device recalls

intelerad IntelePACS software

INTELERAD MEDICAL SYSTEMS INCORPORATED · A cybersecurity vulnerability has been identified in the software that could allow a malicious user to take over an administrator session, delete or modify ima…

Open, ClassifiedSoftware Design ChangeZ-0357-2027
Open FDA's page for this record
Cfres ID
223113
Product res number
Z-0357-2027
Event date initiated
2026-09-10
Event date posted
2026-11-04
Recall status
Open, Classified
Res event number
99971
Product code
LLZ
Product description
intelerad IntelePACS software
Firm FEI number
3003768392
Recalling firm
INTELERAD MEDICAL SYSTEMS INCORPORATED
Address 1
800, Boul. de Maisonneuve Est 800 Boul. de Maisonneuve Est
Address 2
14th floor
City
Montreal
Country
Canada
Root cause description
Software Design Change
Product quantity
560 software platforms
Code info

HIBCC:B228INTELEPACS0/Affected versions: PACS-4-3-4-R20 PACS-4-9-1-R80 PACS-4-12-1-R81 through R90 PACS-4-14-1-R46 through R137 PACS-4-16-1-R50 through R137 PACS-4-17-1-R23 through R117 PACS-4-18-1-R31 through R91 PACS-5-1-1-R17 through R102 PACS-5-2-1-R36 through R93 PACS-5-3-1-R26 through R128 PACS-5-4-1-R15 through R101 PACS-5-5-1-R7 through R89 PACS-5-6-1-R12 through R81 PACS-5-7-1-R2 through R61 PACS-5-8-1-R2 through R42 PACS-5-9-1-R8 through R19

Reason for recall

A cybersecurity vulnerability has been identified in the software that could allow a malicious user to take over an administrator session, delete or modify image or study data, or lock legitimate users out of the system. Additionally, a software issue has been identified that may result in the unintended deletion of image data from an external archive integration.

Action

On September 15, 2026 Intelerad, (a GE HealthCare Company) issued a Urgent Medical Device Correction notification to affected consignees via: email. Intelerad ask consignees to take the following actions: 1. Pending corrections by Intelerad, you can continue to use your system by continuing to follow your organization's standard cybersecurity and access control practices until the corrective update is implemented. 2. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. 3. Please retain this document for your records. 4. Please complete and return the attached acknowledgement form electronically via FMI 88101 Digital Response Form or email, or print, fill out manually, scan, and email to Recall.88101@gehealthcare.com.

Distribution pattern

US: AL FL CO MI CA TX CT AZ NJ IL NC SC MO NH WA KY WY DC NV IN PA AK OR AR KS TN MN MD IA WI HI NY OH VA WV NM GA LA NE UT WA OUS: Australia , Canada, Ireland, New Zealand, South Africa, United Kingdom

K numbers
K192176
Harmonized identifiers (openFDA)
Device name
System, Image Processing, Radiological
Medical specialty description
Radiology
Regulation number
892.2050
Device class
2
K number
K080325, K071972, K140587, K992306, K002780, K974474, K063579, K844317, K143044, K192962, K022692, K103724, K191014, K883476, K080770, K212420, K083658, K130228, K142337, K072952, K033831, K080951, K083327, K000770, K130743, K130770, K033093, K180125, K053043, K123650, K063107, K914366, K092519, K110414, K173456, K243447, K073271, K170250, K180323, K223800 +2,257 more
Registration number
3007301305, 3013665681, 3009196021, 2027467, 3025308066, 3013342228, 3018133052, 3005202320, 3033508292, 3006060685, 3017233423, 3011221136, 3013503781, 3009302839, 3011501361, 3036495186, 9680825, 3007339076, 3033527273, 3014498512, 3006260740, 3013634985, 3030409224, 3003989043, 3016244609, 3013718526, 2134213, 3006894636, 3010119782, 3004753388, 3022973841, 3019252669, 3010493171, 1825014, 3044622018, 3010533307, 3015961056, 3012448806, 3007487007, 3042625805 +603 more
FEI number
3008581796, 3007301305, 3013665681, 3003772586, 3009196021, 2027467, 3002957361, 3025308066, 3013342228, 3018133052, 1000115748, 3005202320, 3033508292, 3006974930, 3004455125, 3006060685, 3017233423, 3011221136, 3013503781, 3009302839, 3011501361, 3001452787, 3036495186, 3007339076, 3033527273, 3014498512, 1000511585, 3006260740, 3013634985, 3030409224, 3003309342, 3003989043, 3016244609, 3013718526, 3006894636, 3010119782, 3004753388, 3022973841, 3019252669, 3010493171 +603 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.