LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303

Luminex Corporation · Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue ma…

Open, ClassifiedError in labelingZ-0359-2027
Open FDA's page for this record
Cfres ID
223172
Product res number
Z-0359-2027
Event date initiated
2026-10-05
Event date posted
2026-11-05
Recall status
Open, Classified
Res event number
100036
Product code
NSU
Product description
LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303
Code info
All lots UDI-DI pending
Firm FEI number
3006028115
Recalling firm
Luminex Corporation
Address 1
4088 Commercial Ave
City
Northbrook
State
IL
Postal code
60062-1829
Additional info contact
Technical Support 877-785-2323 Ext. 9029
Root cause description
Error in labeling
Product quantity
30,168 test cartridges (2,514 Kits)
Distribution pattern
US distribution nationwide. No OUS distribution.
Reason for recall

Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection.

Action

On October 5, 2026, the firm began notifying customers via "Urgent Field Safety Notice - Medical Device Recall" letters. Consignees were informed that the current Instructions for Use for the Gram-Negative Blood Culture Assay do not adequately describe interpretation of certain positive family-level results. Specifically, when a positive family-level only result is obtained ("family-level only result" refers to a positive organism family-level detection without an accompanying genus- or species-level identification), repeat testing with the LIAISON PLEX Gram-Negative Blood Culture Assay is recommended. Absence of this instruction may increase the likelihood that certain reportable detections are not appropriately recognized by the user and may contribute to an incorrect interpretation of test results. Customers were instructed to implement the following interpretation guidance: When a positive family-level only result is obtained, repeat testing with the LIAISON PLEX BCN assay is recommended. DiaSorin is updating the Instructions for Use to include the revised guidance. Revised labeling will be provided following implementation of the labeling revision. Please contact Technical Support Services with any questions or concerns at the following phone numbers:1-877-785-2323 (U.S. and Canada); +1-512-381-4397 (Outside U.S. and Canada); 1-512-219-5114 (Fax) and email:support@luminexcorp.com.

K numbers
K243013
Harmonized identifiers (openFDA)
Device name
Instrumentation For Clinical Multiplex Test Systems
Medical specialty description
Clinical Chemistry
Regulation number
862.2570
Device class
2
K number
K121399, K093128, K143178, K110786, K133302, DEN040012, K082562, K141220, K070597, K080995, K073506, K163652, K133849
Registration number
3009189893, 3021389305, 3003366417, 3013540184, 3002773840, 2243471, 3008632402, 3005269352, 3010461097, 3038610665, 3017358793, 3016611336, 8043379, 1650733, 3005967741, 3007118747, 8020030, 3003673482, 3007208259, 3004530258, 1628664, 3035381641, 2024800, 3017907743, 3014515702, 2183715, 3007420875, 3007361513, 9615056, 3010286073, 3006198300, 3003795116, 3042641161, 9610126, 3010770794, 2032682, 3007143373, 3006028115, 3012259842, 3005370583 +23 more
FEI number
3009189893, 3021389305, 3003366417, 3013540184, 3002773840, 3006461677, 2243471, 3008632402, 3005269352, 3010461097, 3038610665, 3017358793, 3016611336, 3005967741, 3007118747, 3007208259, 3003673482, 3004530258, 1628664, 3035381641, 2024800, 3002806559, 3017907743, 3014515702, 1000186023, 2183715, 3008272466, 3003952975, 3007420875, 3007361513, 3010286073, 3006198300, 3003795116, 3042641161, 3002524000, 3010770794, 3007143373, 3006028115, 3012259842, 3005370583 +23 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.