Device recalls

Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

XTANT Medical Holdings, Inc · Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect be…

Open, ClassifiedProcess controlZ-0846-2024
Open FDA's page for this record
Cfres ID
205344
Product res number
Z-0846-2024
Event date initiated
2023-11-28
Event date posted
2024-01-29
Recall status
Open, Classified
Res event number
93738
Product code
NKB
Product description
Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.
Code info
UDI-DI: 00810076564005, Lot: AP23000143
Firm FEI number
3005031160
Recalling firm
XTANT Medical Holdings, Inc
Address 1
664 Cruiser Ln
City
Belgrade
State
MT
Postal code
59714-9719
Additional info contact
Michele Dolan 406-813-4145
Root cause description
Process control
Action
On 11/28/23 the recall notice was emailed. Customers with questions can call the firm at 1-406-924-5878 from 8 a.m. to 5 p.m. (Mountain Time).
Product quantity
6
Distribution pattern
US distribution in MI.
Reason for recall

Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

K numbers
K221403
Harmonized identifiers (openFDA)
Device name
Thoracolumbosacral Pedicle Screw System
Medical specialty description
Orthopedic
Regulation number
888.3070
Device class
2
K number
K180176, K231744, K163096, K141044, K101971, K180337, K211414, K052054, K203056, K111449, K183117, K222031, K060979, K051275, K223274, K182974, K090320, K100757, K221244, K091782, K111502, K190361, K173882, K201926, K060615, K073267, K113270, K032033, K231069, K070995, K071380, K151648, K170611, K170251, K190600, K073172, K150092, K121342, K221332, K112607 +875 more
Registration number
3008534770, 2027467, 3008754074, 3007922509, 3006563559, 3031261833, 3011964577, 3007279845, 3041915189, 2648623, 3011277306, 3009971621, 3013548554, 3011500362, 9611274, 3010009693, 9615766, 3021006165, 3003639920, 1450662, 3009540749, 3015212339, 3009631939, 3006493760, 3004176895, 9610612, 2133928, 1000200989, 3024999441, 3008850074, 8020776, 1224360, 3007125392, 3004941535, 3015869121, 3021948057, 2133687, 3030412764, 3010863450, 1066368 +375 more
FEI number
3008534770, 2027467, 3008754074, 3007922509, 3006563559, 1000115331, 3031261833, 3011964577, 3007279845, 3041915189, 3001814066, 1000306431, 2648623, 3011277306, 3009971621, 3008647423, 3013548554, 3011500362, 3010009693, 3021006165, 1000221916, 3003639920, 1450662, 3009540749, 3015212339, 3009631939, 3002287318, 3006493760, 3004176895, 3000206619, 1000200989, 3024999441, 3008850074, 3010938185, 3007125392, 3000717902, 3004941535, 3015869121, 3000717687, 3021948057 +375 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.