Ingenia 1.5T CX. Model (REF) Numbers 781262, 781261.

Philips North America · The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the sc…

Open, ClassifiedProcess controlZ-0950-2024
Open FDA's page for this record
Cfres ID
205473
Product res number
Z-0950-2024
Event date initiated
2024-01-03
Event date posted
2024-02-01
Recall status
Open, Classified
Res event number
93787
Product code
LNI
Product description
Ingenia 1.5T CX. Model (REF) Numbers 781262, 781261.
Code info
a. Model No. 781262; UDI-DI 00884838068445. b. Model No. 781261; UDI-DI 00884838068438. All units affected
Firm FEI number
3006648320
Recalling firm
Philips North America
Address 1
222 Jacobs St
City
Cambridge
State
MA
Postal code
02141-2296
Additional info contact
Customer Care Solution Center 800-722-9377
Root cause description
Process control
Product quantity
0 US; 146 OUS
Distribution pattern
Worldwide - US Nationwide distribution.
Reason for recall

The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

Action

On January 4, 2024, the firm began notification of affected customers via URGENT Medical Device Correction letters. Customers were provided with precautions to take when using the device, including inspection of the QBC seal for separation. If the QBC seal is found to be loose, or becomes loose during a patient scan, customers should immediately stop use and contact their local Philips service representative. Customers should also ensure that all users of the device are aware of the issue and retain the letter with affected systems until a solution is installed. Philips will contact customers to schedule a time for a Field Service Engineer (FSE) to visit your site and replace your system's QBC Seal. Philips plans to start implementing corrections in Q3 2024. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377). Revision B of the URGENT Medical Device Correction letter was distributed to the additional US consignees on 24-Mar-2025. It was later identified that UMDC Rev B contained three MR systems with a 70cm bore (Ingenia 1.5T: 781341, Ingenia 3.0T: 781377, Ingenia Ambition X: 782109) which were incorrectly included, as the scope of this field action is MR systems with 60cm bore. The URGENT Medical Device Correction letter was updated to Rev C to remove these systems. Revision C of the UMDC was distributed to the additional US consignees on 09-Apr-2025.

K numbers
K193215
Harmonized identifiers (openFDA)
Device name
System, Nuclear Magnetic Resonance Spectroscopic
Medical specialty description
Radiology
Regulation number
892.1000
Device class
2
K number
K991568, K093044, K930265, K071506, K984167, K962627, K011604, K003575, K970005, K010129
Registration number
3021059265, 3012085143, 3005810333, 2183553, 3004754211, 3012169560, 3018423337, 3009154005, 3023003446, 3012939903, 3002808157, 3017375271, 3005049692, 3010398867, 3010949642, 3031503890, 3035950510, 3004531751, 3008447096, 3025602463, 3013875850, 3014343100, 3032109225, 3009196578, 3029521953, 3012127758, 3008632256, 3003768277, 3026047529, 3016851825, 3005877899, 3009077524, 3013343065, 3015329423, 3015276088, 9616047, 3035371452, 3005326970, 3033527281, 9613936 +12 more
FEI number
3021059265, 3012085143, 3005810333, 1000571342, 2183553, 3004754211, 3012169560, 3018423337, 3009154005, 3023003446, 3012939903, 3002808157, 3017375271, 3005049692, 3010398867, 3010949642, 3031503890, 3035950510, 3004531751, 3008447096, 3025602463, 3013875850, 3014343100, 3032109225, 3009196578, 3029521953, 3012127758, 3008632256, 3003377038, 3026047529, 3016851825, 3005877899, 3009077524, 3013343065, 3015329423, 3015276088, 3035371452, 3005214683, 3005326970, 3033527281 +12 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.