Device recalls

VARIPULSE Bi-Directional Ablation Catheter REF D141201. The Field Catheter is indicated for use in catheter based cardiac electrophysiologi…

Biosense Webster, Inc. · Due to an observed trend of neurovascular events

CompletedLabeling designZ-1205-2025
Open FDA's page for this record
Cfres ID
212216
Product res number
Z-1205-2025
Event date initiated
2025-01-05
Event date posted
2025-02-28
Recall status
Completed
Res event number
96170
Product code
QZI
Firm FEI number
3013300026
Recalling firm
Biosense Webster, Inc.
Address 1
31 Technology Dr Ste 200 31 Technology Dr
Address 2
Ste 200
City
Irvine
State
CA
Postal code
92618-2302
Additional info contact
909-839-8500
Reason for recall
Due to an observed trend of neurovascular events
Root cause description
Labeling design
Product quantity
497 total catheter - 378 US and 119 OUS
Product description

VARIPULSE Bi-Directional Ablation Catheter REF D141201. The Field Catheter is indicated for use in catheter based cardiac electrophysiological mapping (stimulating and recording) and, when used for TRUPULSE Generator, for the treatment of drug refractory recurrent symptomatic paroxysmal atrial fibrillation. The catheter provides location information when used with the CARTO 3 System.

Code info

Product Number: D141201-12 UDI-DI code: 10846835025460 Batch Numbers: 31483574L 31483560L 31483581L 31483542L 31487313L 31483552L 31483544L 31487311L 31483550L 31483558L 31483575L 31483556L 31487312L 31483553L 31483565L 31483576L 31483547L 31487320L 31483549L 31483561L 31483548L 31483564L 31483541L 31483546L 31483562L 31483566L 31483583L 31483567L

Action

On 01/05/2025, the firm called and emailed customers to inform them to pause the US External Evaluations and all Varipulse cases due an observed trend in neurovascular events associated with the use of the VariPulse Bi-Directional Ablation Catheters distributed between December 2024 and January 2025. For U.S. Customers, Biosense Webster/JNJ Medtech EP is pausing the US External Evaluation and all US Varipulse cases. Healthcare providers who have treated patients using the products prior to the pause should continue to follow those patients according to their standard of care. For any questions, contact Biosense Webster/JNJ MEDTECH EP representatives On 01/08/2025, the firm released the information on the firm's parent company's website https://www.jnj.com/media-center/press-releases/johnson-johnson-statement-on-varipulse-u-s-external-evaluation

Distribution pattern

Worldwide distribution; US states of AR, CA, CO, KS, LA, MA, MO, NY, OH and TX. EMEA, New Zealand, Saudi Arabia, Oman, United Ecuador, Panama, South Africa, China Austria Belgium Czech Republic France Germany Hungary Ireland Israel Italy UK Sweden Norway Denmark Finland Luxembourg Netherlands Poland Portugal Slovenia Spain Switzerland Serbia United Arab Emirates Croatia Canada South Korea Japan

PMA numbers
P240006
Harmonized identifiers (openFDA)
Device name
Percutaneous Cardiac Ablation Catheter For Treatment Of Atrial Fibrillation With Irreversible Electroporation
Medical specialty description
Unknown
Device class
3
PMA number
P240006, P250022, P240013, P230017, P240044, P230030
Registration number
3009504230, 3000279201, 3001592626, 3007603855, 3002648230, 1000121056, 2124215, 3008093809, 2182269, 3012263546, 3013403214, 2183744, 2024024, 3008452825, 3008328679, 3004859241, 2032112, 3007648309, 3012693387, 3005334138, 3005012805, 3004016520, 3017060084, 3009243923, 3012520654, 9681260, 3007029079, 9680579, 3013300026, 1319639, 1721686, 2184149, 2182208, 3004105270, 1721983, 3016245399, 1928237, 3012179728, 3004593495, 3012260771 +5 more
FEI number
3009504230, 3000279201, 3001592626, 3007603855, 3002648230, 1000121056, 3002806565, 2124215, 3008093809, 2182269, 3012263546, 3013403214, 2183744, 2024024, 3008452825, 3002806434, 3008328679, 3004859241, 3007648309, 3012693387, 3005334138, 3005012805, 3004016520, 3017060084, 3009243923, 3012520654, 3007029079, 1000064530, 3013300026, 1721686, 3004105270, 2182208, 3016245399, 1928237, 3012179728, 1000139754, 3004593495, 1000138054, 3003076831, 3003248396 +5 more