Device recalls

human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.

HUMAN MED AG · The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather th…

Open, ClassifiedProcess controlZ-1251-2025
Open FDA's page for this record
Cfres ID
212112
Product res number
Z-1251-2025
Event date initiated
2024-10-18
Event date posted
2025-02-28
Recall status
Open, Classified
Res event number
96137
Product code
QPB
Product description
human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.
Code info
Lot number 24-20198, UDI (01)04260170881014(11)240611(17)270611(10)24-20198; and Lot number 23-19612, UDI (01)04260170881014(11)240109(17)270109(10)23-19612.
Firm FEI number
3003587189
Recalling firm
HUMAN MED AG
Address 1
Wilhelm-Hennemann-Str. 9 9 wilhelm-hennemann-str.
City
Schwerin
Country
Germany
Root cause description
Process control
Product quantity
775 devices
Distribution pattern
US Nationwide distribution in the state of FL.
Reason for recall

The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet.

Action

The recalling firm issued a Field Safety Notice to Distributors dated 10/18/2025 via email explaining the reason for the Field Safety Notice, the potential hazard, and informing the consignee to discontinue delivering the affected products to customers and users and to dispose of the product if still in their warehouse. A Field Safety Notice was enclosed to be issued to the consignee's users and customers who received the affected product from the Distributor. The User and Customer letter dated 10/17/2024 was similar to the Distributor letter except provided instructions to check the spray pattern as instructed in the user manuals prior to using the cannula. If the pattern was not in a flat, fan-shaped pattern, they are not to use the cannula and it should be sorted out and disposed of properly. A confirmation form was enclosed to confirm the cannula has been correctly disposed of and the form was to be sent to the distributor or to the email address provided.

K numbers
K082025
Harmonized identifiers (openFDA)
Device name
System, Suction, Lipoplasty For Removal
Medical specialty description
General, Plastic Surgery
Regulation number
878.5040
Device class
2
K number
K244026, K253396, K063336, K101413, K093067, K252495, K244050, K241835, K082025, K980763, K990602, K981527, K052183, K190551, K190555, K981922, K261710, K113795, K161722, K251944, K981215, K012042, K254135, K220318, K150134, K001803, K012236, K042261, K031881, K121218, K004005, K063235, K192694, K240188, K081039, K200168, K041058, K981172, K992855, K170329 +20 more
Registration number
3007593903, 3002807523, 3007697975, 3011423170, 3018094310, 3038613815, 1222928, 2027804, 2030624, 3030230686, 3022978896, 3009039068, 9680741, 2032809, 2022994, 3008837339, 3009103001, 3004112621, 1000655158, 3004878380, 1058584, 3009420598, 3003587189, 3038715747, 1220246, 3003755939, 2032098, 3008023310, 1047843, 2020601, 1118193, 3044777117, 1045254, 1645337, 3007279404, 3007829657, 1313525, 3011533696, 3014903258, 3005670221 +10 more
FEI number
3007593903, 3002807523, 1000219637, 3007697975, 3018094310, 3038613815, 3004106598, 3030230686, 3022978896, 3009039068, 1000125732, 3008837339, 3009103001, 1000118360, 1000125147, 2022994, 3004112621, 1000655158, 3004878380, 3009420598, 3004169154, 1000120613, 3003587189, 3038715747, 1220246, 3003755939, 3008023310, 1047843, 3002807634, 2020601, 1118193, 3003058448, 1317628, 3044777117, 3007279404, 1045254, 3007829657, 3000206435, 3011533696, 3014903258 +10 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.