MIM software; System, Image Processing, Radiological

MIM Software Inc · In situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) wit…

Open, ClassifiedSoftware designZ-1267-2025
Open FDA's page for this record
Cfres ID
212540
Product res number
Z-1267-2025
Event date initiated
2025-02-19
Event date posted
2025-03-04
Recall status
Open, Classified
Res event number
96314
Product code
LLZ
Product description
MIM software; System, Image Processing, Radiological
Firm FEI number
3004363352
Recalling firm
MIM Software Inc
Address 1
25800 Science Park Dr Ste 180 25800 Science Park Dr
Address 2
Ste 180
City
Cleveland
State
OH
Postal code
44122-7311
Additional info contact
216-455-0600
Root cause description
Software design
Product quantity
352 units
Code info

Software versions: 1. MIM Anyware 1.1, UDI/DI Not Applicable 2. MIM 7.2.0, UDI/DI 00850009343192; 3. MIM 7.2.1, UDI/DI 00850009343192; 4. MIM 7.2.2, UDI/DI 00850009343192; 5. MIM 7.2.3, UDI/DI 00850009343192; 6. MIM 7.2.4, UDI/DI 00850009343192; 7. MIM 7.2.5, UDI/DI 00850009343192; 8. MIM 7.2.6, UDI/DI 00850009343192

Reason for recall

In situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software versions 7.2.0 through 7.2.6.could result.

Action

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/19/2025 disseminated using a traceable means. The notice explained the issue, potential risk to the patient, and provided the following actions to be taken to use the device: "You can continue to use your device if you are not using the SUV tool or if you access a non-impacted version of MIM when using the SUV tool. If using a non-default workflow and fusing a PET or SPECT image with another image with a different FOV, do not incorporate the Maximum SUV value into assessment of disease status, progression or response to treatment. NOTE: MIM s default workflows do not fuse images with differing FOV, and the Maximum SUV values reported are correct. If you are using the SUV tool and are unsure if you are affected by this issue, please contact your MIM Site Development Manager, your Distributor for MIM Software, or MIM Customer Support." "Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions."

Distribution pattern

Worldwide distribution - United States Nationwide and the countries of Australia, Belgium, Bolivia, Canada, China, Croatia, Czech Republic, Denmark, France, Germany, Greece, Israel, Italy, Japan, Mexico, Netherlands, Poland, Russia, Spain, Sweden, United Kingdom, Finland, Hong Kong.

K numbers
K071964
Harmonized identifiers (openFDA)
Device name
System, Image Processing, Radiological
Medical specialty description
Radiology
Regulation number
892.2050
Device class
2
K number
K061696, K915305, K110573, K243697, K220727, K992925, K172477, K253111, K212981, K953616, K131246, K182624, K962153, K042634, K973983, K962589, K063006, K041990, K043146, K140139, K000770, K071331, K897000, K925963, K171314, K963697, K131822, K141570, K880169, K182904, K152602, K983821, K182230, K033849, K002942, K042507, K843708, K221984, K130395, K061029 +2,255 more
Registration number
3001052305, 3005406110, 3017545364, 3005172103, 9617978, 3004706313, 3008487708, 1226544, 3009341091, 3044776706, 3010656400, 3038189169, 3007362683, 3007993775, 1061158, 3014232555, 3003757027, 3014991167, 9617592, 3013634985, 3023217364, 1054415, 3013535906, 3026047529, 3019878658, 3005225421, 3030655434, 3009156722, 3010445368, 8020825, 3011676026, 3008364043, 3045521711, 3014635169, 2954921, 9615439, 3011015597, 3003989064, 3035371452, 3019544554 +603 more
FEI number
3001052305, 3005406110, 3017545364, 3005172103, 3004706313, 3008487708, 3009341091, 3044776706, 3002808188, 3010656400, 3038189169, 3007362683, 3007993775, 3014232555, 3003757027, 3014991167, 3013634985, 3023217364, 3013535906, 3026047529, 3019878658, 3005225421, 3030655434, 3009156722, 3010445368, 3011676026, 3008364043, 3045521711, 3014635169, 3011015597, 3003989064, 3035371452, 3019544554, 3008548323, 3014269683, 3002821504, 2126677, 3002619595, 3010684280, 3003861317 +603 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.