Device recalls

Vios Monitoring System Bedside Monitor Model BSM2050

Murata Vios, Inc. · Batteries may deplete over time while not in use if devices are stored without being plugged in to wall outlet. Depleted batteries may lead to tablet connectiv…

Open, ClassifiedLabeling designZ-1292-2025
Open FDA's page for this record
Cfres ID
212419
Product res number
Z-1292-2025
Event date initiated
2024-08-09
Event date posted
2025-03-06
Recall status
Open, Classified
Res event number
96252
Product code
DRT
Product description
Vios Monitoring System Bedside Monitor Model BSM2050
Code info
1.56.01 UDI: +B595BSM20500/$$7V1.56.01/16D20230630S 1.57.20 UDI: +B595BSM20500/$$7V1.57.20/16D20231129Y Software Versions 1.56.01 and 1.57.20
Firm FEI number
3015328411
Recalling firm
Murata Vios, Inc.
Address 1
700 Commerce Dr Ste 190 700 Commerce Dr
Address 2
Ste 190
City
Woodbury
State
MN
Postal code
55125-9239
Additional info contact
Customer Service Team 651-764-8467
Root cause description
Labeling design
Product quantity
50
Distribution pattern
US Nationwide distribution in the states of Florida, Louisiana, Maryland, Montana, New York, Ohio, and Texas.
Reason for recall

Batteries may deplete over time while not in use if devices are stored without being plugged in to wall outlet. Depleted batteries may lead to tablet connectivity issues.

Action

The firm sent a single page of instructions to customers on August 9, 2024 via email. Customers were informed that tablets should remain plugged into a wall outlet when the tablet is not in use. The tablet battery should not be allowed to fully deplete at any time during patient monitoring or when not in use.

K numbers
K172586
Harmonized identifiers (openFDA)
Device name
Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm)
Medical specialty description
Cardiovascular
Regulation number
870.2300
Device class
2
K number
K902652, K801761, K761076, K863494, K840525, K851335, K802204, K944262, K842236, K813444, K832131, K910870, K790454, K830795, K151781, K915624, K844911, K842462, K821942, K926298, K883069, K110987, K901031, K202464, K820540, K840521, K841499, K840524, K844348, K820671, K892794, K832018, K821368, K822692, K972789, K832021, K981378, K790552, K831752, K851916 +301 more
Registration number
3003263092, 3015282442, 3008729547, 1319390, 3008716327, 3010703925, 3003589956, 9611031, 3009499478, 3016852448, 1054713, 3007048291, 3016592682, 3010390468, 3016674095, 3015181391, 3014541700, 3009156722, 3007413079, 3016618143, 3012359877, 3013679558, 9710654, 3013826848, 3026751377, 3014848734, 3004977335, 9610816, 2221819, 3010698831, 3003971136, 3004415095, 3008483389, 2080783, 3003800159, 3007305624, 3007603826, 3005521864, 1527715, 3000126629 +64 more
FEI number
3003263092, 3015282442, 3008729547, 1319390, 3008716327, 3010703925, 3003589956, 3009499478, 3016852448, 3007048291, 3016592682, 3010390468, 3000957797, 3016674095, 3015181391, 3014541700, 3009156722, 3007413079, 3016618143, 3012359877, 3013679558, 3013826848, 3026751377, 3014848734, 3004977335, 2221819, 3010698831, 3003971136, 3000215024, 3004415095, 3008483389, 3002807739, 2080783, 3003800159, 3007305624, 3007603826, 3005521864, 1527715, 3000126629, 3002807720 +64 more