Device recalls

GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed

GE Medical Systems, LLC · GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the…

Open, ClassifiedProcess change controlZ-1301-2025
Open FDA's page for this record
Cfres ID
212665
Product res number
Z-1301-2025
Event date initiated
2025-02-18
Event date posted
2025-03-07
Recall status
Open, Classified
Res event number
96370
Product code
JAK
Product description
GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed
Code info
System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN
Firm FEI number
2126677
Recalling firm
GE Medical Systems, LLC
Address 1
3000 N Grandview Blvd
City
Waukesha
State
WI
Postal code
53188-1615
Additional info contact
GE HealthCare Service 800-437-1171
Root cause description
Process change control
Product quantity
1 unit
Distribution pattern
Worldwide distribution.
Reason for recall

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Action

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/18/2025 via letter delivered using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your system. Please monitor the floor near the base of the gantry for any accumulation of blue fluid. If blue fluid is observed, wear nitrile or rubber gloves, clean up the fluid by absorbing it with a noncombustible absorbent material, and dispose of the absorbent material in a suitable container for disposal. Please contact your local GE HealthCare service representative if any blue fluid is observed. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please complete and return the attached acknowledgement form to recall.25504@gehealthcare.com."

K numbers
K213715
Harmonized identifiers (openFDA)
Device name
System, X-Ray, Tomography, Computed
Medical specialty description
Radiology
Regulation number
892.1750
Device class
2
K number
K151376, K183204, K896456, K252217, K974344, K840860, K113287, K913485, K003736, K961464, K894542, K032243, K041521, K820688, K192763, K250970, K023687, K213096, K003639, K023332, K991553, K803181, K012238, K243523, K772291, K173244, K171013, K211180, K173821, K192188, K974455, K223424, K910375, K022239, K241200, K241440, K091103, K203699, K260391, K201586 +787 more
Registration number
3007225492, 3027338963, 3018094310, 9617978, 3015223550, 3018423337, 3002807636, 3015169361, 3010703925, 3003768277, 3029521953, 1066019, 3017375271, 3009154005, 3004962788, 9616567, 3004091527, 3012957335, 3010398867, 3016965352, 2247992, 3003575350, 3002808157, 1054713, 3006209933, 3004857110, 3008114965, 3026047529, 3043512431, 3003477135, 3008284470, 3009196578, 3031503890, 1000246715, 3019216, 3001722928, 3013679558, 3004753774, 3005156333, 3011015597 +125 more
FEI number
3007225492, 3027338963, 3018094310, 3015223550, 3018423337, 3002807636, 3015169361, 3010703925, 3002807903, 3008581796, 3029521953, 3003079232, 1000139922, 3017375271, 3009154005, 3004962788, 3004091527, 3012957335, 3010398867, 3016965352, 3003575350, 3002808157, 3002808529, 3002808382, 1000123143, 3006209933, 3004857110, 3008114965, 3026047529, 3043512431, 3003477135, 3008284470, 3009196578, 3000206435, 3031503890, 1000246715, 3019216, 3001722928, 3013679558, 3004753774 +125 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.