Device recalls

Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754

Terumo Cardiovascular Systems Corporation · During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. I…

Open, ClassifiedSoftware designZ-1492-2025
Open FDA's page for this record
Cfres ID
212566
Product res number
Z-1492-2025
Event date initiated
2025-03-03
Event date posted
2025-03-31
Recall status
Open, Classified
Res event number
96331
Product code
DRY
Product description
Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754
Code info
UDI/DI 00886799001882, Serial Number range: H0050095-H0050110, H0050112-H0050116, H0050118-H0050124, H0050126, H0050131-H0050134, H0050136, H0050140-H0050142
Firm FEI number
1828100
Recalling firm
Terumo Cardiovascular Systems Corporation
Address 1
6200 Jackson Rd
City
Ann Arbor
State
MI
Postal code
48103-9586
Additional info contact
customer service 800-521-2818
Root cause description
Software design
Product quantity
37 units
Distribution pattern
US: IL, TX, NY, MI, FL, NV, and Australia, New Zealand
Reason for recall

During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.

Action

Terumo Cardiovascular issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 03/03/2025 via email. The notice explained the reason for the removal, potential hazard, and requested the following: 1. Review this Medical Device Correction and assure that all users have received notice of this issue. 2. Confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. 3. Terumo CVS will contact users to discuss removal and correction activities. Note: Terumo CVS recommends that users continue using CDI OneView System H/S Probes while waiting for this removal

K numbers
K234065
Harmonized identifiers (openFDA)
Device name
Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass
Medical specialty description
Cardiovascular
Regulation number
870.4330
Device class
2
K number
K852124, K874839, K881006, K133658, K915255, K871686, K900459, K202557, K972962, K862069, K903729, K234065, K931968, K955379, K900682, K854357, K012743, K862953, K062468, K071725, K110957, K123039, K202154, K973472, K843538, K133157, K091908, K103168, K941922, K861277, K173591, K955380, K980732, K181923, K072131, K871221, K915265, K910421, K831225, K821159 +7 more
Registration number
1124841, 3019807891, 3000126629, 9611109, 1118880, 3002807090, 1218950, 3006073153, 1319639, 3016618143, 2184009, 3044156918, 1928237, 1828100, 2245590, 2029275, 3011560054, 3002807561, 9617601, 2011171, 9680841
FEI number
3019807891, 3000204839, 3002808267, 3000126629, 1000306431, 1118880, 3005426550, 1218950, 1000116158, 3006073153, 3016618143, 3006003617, 3044156918, 1928237, 1828100, 2245590, 1000138054, 3011560054, 3002534070, 3002807561, 2011171
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.