Device recalls

BD Connecta BD Luer-Lok 360, REF 394910 UDI-DI code: 00382903949106 Connecta Plus 1 and Connecta Plus 3 2-way and 3-way Stopcocks are contr…

BD SWITZERLAND SARL · Due to specific lots of luer-lok devices were shipped to the U.S. market with an Outside US Instruction for Use (IFU)

Open, ClassifiedLabeling Change ControlZ-1509-2025
Open FDA's page for this record
Cfres ID
212777
Product res number
Z-1509-2025
Event date initiated
2025-02-20
Event date posted
2025-04-02
Recall status
Open, Classified
Res event number
96399
Product code
FMG
Code info
Catalog Number: 394910 UDI-DI code: 00382903949106 Lot Numbers: 4277968 4277973 4282451 4282461 4289801 4290582 4292904 4295848 4305108 4306700
Firm FEI number
3012798826
Recalling firm
BD SWITZERLAND SARL
Address 1
17 terre bonne park a4 route de crassier
City
Eysins
Country
Switzerland
Additional info contact
Krista DiPaula 410-598-9328
Reason for recall
Due to specific lots of luer-lok devices were shipped to the U.S. market with an Outside US Instruction for Use (IFU)
Root cause description
Labeling Change Control
Product quantity
500 units
Distribution pattern
U.S.: AK, AL, AR, AZ, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NV, OH, OK, OR, PA, SC, TN. TX, UT, VA, WA, WV O.U.S.: N/A
Product description

BD Connecta BD Luer-Lok 360, REF 394910 UDI-DI code: 00382903949106 Connecta Plus 1 and Connecta Plus 3 2-way and 3-way Stopcocks are control valves for use in IV. therapy and hemodynamic pressure monitoring.

Action

On 02/20/2025, the firm sent via FedEx and/or email an "URGENT: Medical Device Product Recall MDS-25-5241" letter informing customers that BD confirmed that specific lots of BD Connecta BD Luer-Lok 360 were shipped to the U.S. market with Instruction for Use (IFU) labeling meant for Outside U.S. market. Customers are instructed to: 1. Locate any unused affected BD Connecta" BD Luer-Lok" 360, as listed in Table 1. 2. Remove the IFU from each shelf-box and destroy it. (OUS IFU version, Document number 500037734). 3. Insert a copy of Attachment 1 BD Connecta IFU Document Number: 500081278 (US IFU version). 4. Circulate this notice to all those who need to be aware within their organization or to any organization where the potentially affected products have been transferred. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not customers have any of the impacted material so that BD may acknowledge their receipt of this notification. Report any questions or complaints experienced with the use of this product to BD via the North American Regional Complaint Center: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT Email: productcomplaints@bd.com or bdrc55@bd.com

K numbers
K974083
Harmonized identifiers (openFDA)
Device name
Stopcock, I.V. Set
Medical specialty description
General Hospital
Regulation number
880.5440
Device class
2
K number
K800862, K790010, K022895, K870261, K881805, K070143, K954984, K223175, K141254, K190539, K854348, K792126, K061143, K982368, K961982, K010217, K850280, K823373, K792125, K032393, K053405, K242255, K172266, K080077, K810375, K870324, K812927, K851986, K800864, K792128, K964435, K210516, K123084, K862691, K211204, K130245, K082106, K884435, K792124, K800284 +49 more
Registration number
9616088, 9617594, 1061124, 1055236, 3006425876, 9616567, 3004111573, 3033589330, 1526863, 3013764800, 1518293, 9680794, 1625425, 2523676, 9613418, 3007738736, 3012798826, 2025816, 3032916632, 1018470, 2024024, 9614279, 9610847, 3003704957, 3010421104, 1060680, 3015173212, 3010432890, 1319639, 3002907620, 1423507, 2022435, 3045322464, 3007017524, 3012261967, 8020616, 1649914, 1713468, 3012809732, 1047429 +47 more
FEI number
3002806588, 1000208624, 1000120555, 3006425876, 3002806901, 3001044237, 3003015552, 3004111573, 3033589330, 3001236905, 3003081151, 3013764800, 1526863, 1518293, 1625425, 3002806476, 1000138054, 3003076831, 2523676, 3007738736, 3012798826, 2025816, 3032916632, 3002879021, 1018470, 3003767881, 3003737899, 2024024, 3000215273, 3003040625, 3003058448, 3003704957, 3010421104, 3002807391, 3015173212, 3010432890, 3002907620, 2022435, 3045322464, 3002902735 +47 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.