Zenition 50. Product Code 718096. The devices are used for radiological guidance and visualization during diagnostic, interventional and su…

Philips North America Llc · A wireless foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation.

Open, ClassifiedLabeling designZ-1580-2025
Open FDA's page for this record
Cfres ID
213227
Product res number
Z-1580-2025
Event date initiated
2025-02-25
Event date posted
2025-04-21
Recall status
Open, Classified
Res event number
96593
Product code
OWB
Firm FEI number
3016618143
Recalling firm
Philips North America Llc
Address 1
222 Jacobs St
City
Cambridge
State
MA
Postal code
02141-2289
Reason for recall
A wireless foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation.
Root cause description
Labeling design
Distribution pattern
US
Product description

Zenition 50. Product Code 718096. The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the devices.

Code info

Product UDI: (01)00884838091535(21). 459801442981, Wireless foot switch set 3P (WFSS 3P) 459801238231, Wireless foot switch 3P, 459801257861, WFS Base Station 459800627394, Wireless foot switch set 3P 459800415535, Wireless foot switch 3P, 459800415542, WFS Base Station

Action

Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide the actions that should be taken in order to prevent risks to patients. Philips will contact customers to schedule a visit to inspect the foot switch and provide an addendum to the Instructions for Use (IFU) of the system detailing the appropriate cleaning and maintenance instructions. Philips proposes sending the Electronic Product Radiation Defect notification to customers immediately upon FDA approval of the corrective action plan, and completion of the remaining activities within 4 months of FDA approval of the corrective action plan. This corrective action does not require any customer transportation reimbursement. The following statement is included in the customer communication: The manufacturer will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. Draft of letter provided. Philips commits to notifying the FDA on a monthly basis providing progress with this field correction.

K numbers
K183101
Harmonized identifiers (openFDA)
Device name
Interventional Fluoroscopic X-Ray System
Medical specialty description
Radiology
Regulation number
892.1650
Device class
2
K number
K221230, K151598, K103416, K171565, K043076, K253584, K033486, K083545, K051064, K234067, K070123, K150967, K212383, K200360, K213768, K993037, K220311, K220409, K041931, K171755, K122457, K212557, K232344, K200074, K151904, K220342, K252503, K243735, K231701, K101868, K024200, K233107, K250241, K142740, K233230, K132970, K213479, K121772, K230949, K082243 +285 more
Registration number
3033525987, 1720753, 3035950510, 3009963139, 1056553, 3002466018, 3030089787, 3021027284, 8043933, 3014328942, 1054713, 3005172103, 3011713200, 3018077344, 3006544299, 3010703925, 1835251, 3036255956, 3014343100, 3008284470, 3008496839, 3011347852, 2243057, 3010162973, 3010398867, 3014134519, 3005706667, 3004977335, 3003768277, 3003202425, 3049422862, 3011416394, 3004785967, 3012631246, 3043048939, 3018596740, 3016619943, 3003477135, 3012911444, 3021060474 +86 more
FEI number
3033525987, 1720753, 3035950510, 3009963139, 3002466018, 3030089787, 3000209794, 3021027284, 3014328942, 3005172103, 3011713200, 3018077344, 3006544299, 3010703925, 3036255956, 3014343100, 3008284470, 1000208827, 3008496839, 3011347852, 2243057, 3010398867, 3014134519, 3005706667, 3004977335, 3003202425, 3002807880, 3049422862, 3011416394, 3004785967, 3012631246, 3002808139, 3043048939, 3018596740, 3016619943, 3000205591, 3003477135, 3002980729, 3012911444, 3021060474 +86 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.