Device recalls

Zenition 70. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile…

Philips Medical Systems Nederland B.V. · Philips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch.

Open, ClassifiedSoftware designZ-1583-2025
Open FDA's page for this record
Cfres ID
213230
Product res number
Z-1583-2025
Event date initiated
2023-07-20
Event date posted
2025-04-21
Recall status
Open, Classified
Res event number
96598
Product code
OWB
Code info
Zenition 70
Firm FEI number
3000976525
Recalling firm
Philips Medical Systems Nederland B.V.
Address 1
High Tech Campus 36 36 high tech campus
City
Eindhoven
Country
Netherlands
Additional info contact
Dusty Leppert, CHP 978-228-0190
Reason for recall
Philips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch.
Root cause description
Software design
Product quantity
12 systems
Distribution pattern
US Nationwide distribution
Product description

Zenition 70. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.

Action

Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide the following actions that should be taken to prevent risks to patients: Keep the Electronic Product Radiation Correction letter with the documentation of the system. Where possible, place the Wireless Foot Switch within a 3 meter range from the C-arm stand to minimize as much as possible the potential delay of 72 milliseconds. Circulate this Electronic Product Radiation Correction letter to all users of the system so that they are aware of the issue. Philips will contact customers to schedule a visit to arrange for a Field Service Engineer visit to update the Software free of charge. (Philips reference 2024-IGT-PUN-007).

K numbers
K183040, K212813
Harmonized identifiers (openFDA)
Device name
Interventional Fluoroscopic X-Ray System
Medical specialty description
Radiology
Regulation number
892.1650
Device class
2
K number
K221230, K151598, K103416, K171565, K043076, K253584, K033486, K083545, K051064, K234067, K070123, K150967, K212383, K200360, K213768, K993037, K220311, K220409, K041931, K171755, K122457, K212557, K232344, K200074, K151904, K220342, K252503, K243735, K231701, K101868, K024200, K233107, K250241, K142740, K233230, K132970, K213479, K121772, K230949, K082243 +285 more
Registration number
3033525987, 1720753, 3035950510, 3009963139, 1056553, 3002466018, 3030089787, 3021027284, 8043933, 3014328942, 1054713, 3005172103, 3011713200, 3018077344, 3006544299, 3010703925, 1835251, 3036255956, 3014343100, 3008284470, 3008496839, 3011347852, 2243057, 3010162973, 3010398867, 3014134519, 3005706667, 3004977335, 3003768277, 3003202425, 3049422862, 3011416394, 3004785967, 3012631246, 3043048939, 3018596740, 3016619943, 3003477135, 3012911444, 3021060474 +86 more
FEI number
3033525987, 1720753, 3035950510, 3009963139, 3002466018, 3030089787, 3000209794, 3021027284, 3014328942, 3005172103, 3011713200, 3018077344, 3006544299, 3010703925, 3036255956, 3014343100, 3008284470, 1000208827, 3008496839, 3011347852, 2243057, 3010398867, 3014134519, 3005706667, 3004977335, 3003202425, 3002807880, 3049422862, 3011416394, 3004785967, 3012631246, 3002808139, 3043048939, 3018596740, 3016619943, 3000205591, 3003477135, 3002980729, 3012911444, 3021060474 +86 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.