Device recalls

Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +3 Neck, Model Number 802602203; Prosthesis, Hip, Semi-Constrained, Metal/Cerami…

Zimmer, Inc. · The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome…

Open, ClassifiedUnder Investigation by firmZ-1646-2025
Open FDA's page for this record
Cfres ID
213137
Product res number
Z-1646-2025
Event date initiated
2025-03-24
Event date posted
2025-04-21
Recall status
Open, Classified
Res event number
96493
Product code
LZO
Code info
GTIN 00887868376337, All lots distributed with IFU 87-6204-051-99 Rev D or prior
Firm FEI number
1000220733
Recalling firm
Zimmer, Inc.
Address 1
1800 W Center St
City
Warsaw
State
IN
Postal code
46580-2304
Additional info contact
411 Technical Services 574-267-6131
Root cause description
Under Investigation by firm
Product quantity
215 units
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.
Product description

Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +3 Neck, Model Number 802602203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented

Reason for recall

The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

Action

Zimmer Biomet issued an URGENT MEDICAL DEVICE RECALL CORRECTION - PHASE 1 notice to its consignees on 03/24/2025 via email and courier. The notice explains the issue, potential risk, and requested the following: Distributors: Your Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 3. Return the Additional Accounts form to CorporateQuality.PostMarket@zimmerbiomet.com. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. 4. Retain a copy of your Inventory Return Certification and product return forms for your records in the event of a compliance audit of your facility. Risk Manager Responsibilities: 1. Review this correction notice and ensure that affected personnel, including surgeons, are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with this notice ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgeme

K numbers
K181171
Harmonized identifiers (openFDA)
Device name
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Medical specialty description
Orthopedic
Regulation number
888.3353
Device class
2
K number
K920219, K051588, K935193, K062684, K202768, K934457, K190535, K915184, K120211, K990261, K060577, K191399, K120362, K242249, K892644, K182048, K182705, K181744, K935452, K033944, K240391, K945091, K965002, K161830, K192416, K011533, K192352, K971409, K971497, K991226, K093944, K964431, K934694, K100412, K211891, K915309, K945094, K210678, K992144, K121665 +495 more
Registration number
1000577893, 3009888740, 9681642, 3003639920, 3010155661, 3013450937, 2183744, 1822565, 3016438761, 3043543260, 3009882462, 1057425, 3036756245, 3005913274, 3007334927, 8043792, 3013413393, 3014763043, 1450662, 2245304, 3005333358, 1020279, 3005751028, 3003998208, 9610726, 9610612, 1722511, 1721676, 1836357, 3010047454, 3008668801, 3007353762, 3014315560, 3009732568, 3002498892, 3023852420, 3012275107, 3004371426, 3020282716, 3018052045 +238 more
FEI number
1000577893, 1000220733, 3009888740, 3003639920, 3013450937, 3010155661, 3000206435, 2183744, 1000306303, 3003015552, 3016438761, 3043543260, 3003070421, 1000149660, 3009882462, 3002806315, 3036756245, 3005913274, 3007334927, 1020279, 3013413393, 3014763043, 1450662, 2245304, 3005333358, 3005751028, 3003998208, 1721676, 3010047454, 3008668801, 3007353762, 3014315560, 3003051987, 3009732568, 3002498892, 3023852420, 1000395711, 3002808293, 3012275107, 3002807113 +238 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.