Device recalls

Automated Impella Controller (AIC); Product Code: 0042-0000-US;

Abiomed, Inc. · A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates hav…

Open, ClassifiedUnder Investigation by firmZ-2211-2026
Open FDA's page for this record
Cfres ID
219755
Product res number
Z-2211-2026
Event date initiated
2026-04-20
Event date posted
2026-05-22
Recall status
Open, Classified
Res event number
98770
Product code
OZD
Product description
Automated Impella Controller (AIC); Product Code: 0042-0000-US;
Firm FEI number
1220648
Recalling firm
Abiomed, Inc.
Address 1
22 Cherry Hill Dr
City
Danvers
State
MA
Postal code
01923-2575
Additional info contact
Morgaine Johnson 612-437-5734
Root cause description
Under Investigation by firm
Product quantity
91 units
Code info

Product Code: 0042-0000-US; UDI-DI: 00813502010022; Serial Numbers: IC2449, IC2451, IC2452, IC2453, IC2459, IC2460, IC1096, IC1142, IC1202, IC1203, IC1224, IC1231, IC1238, IC1284, IC1285, IC1390, IC1545, IC1577, IC1665, IC1708, IC1741, IC1852, IC1853, IC2029, IC2071, IC2154, IC2277, IC2278, IC2314, IC2315, IC2368, IC2369, IC2370, IC2442, IC2443, IC2444, IC2461, IC2471, IC2476, IC2477, IC2478, IC2479, IC2575, IC2989, IC3127, IC3128, IC3129, IC3130, IC3166, IC3167, IC3213, IC3223, IC3306, IC3307, IC3308, IC3309, IC3342, IC3398, IC3475, IC3656, IC3658, IC3663, IC3664, IC3665, IC3666, IC3674, IC3675, IC3719, IC3720, IC3776, IC3812, IC3887, IC3888, IC4045, IC4060, IC1301, IC1544, IC1572, IC2686, IC2687, IC2848, IC3174, IC3269, IC3272, IC3277, IC3278, IC3281, IC3282, IC3284, IC3294, IC3783;

Reason for recall

A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.

Action

On April 20, 2026 URGENT MEDICAL DEVICE RECALL (REMOVAL) letters were sent to customers via 1-day direct mail (US). OUS notification process will be commensurate with local requirements. Actions to be taken: 1. For the devices identified, the Abiomed servicing team will contact you to coordinate the return of the device(s) to implement necessary changes. To ensure continuity of care, hospital inventory can continue to be used. 2. Upon contact from Abiomed s field servicing team, please work with them to return the identified device(s) for the change to be implemented. Refer to Attachment 2 for list of impacted AIC serial numbers. 3. Forward to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). 4. If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.

Distribution pattern

Worldwide - US Nationwide distribution in the states of CA, CT, FL, LA, MA, MI, MO, MS, ND, NY, OH, PA, TX, WY and the countries of Brazil, Canada, Colombia, Hong Kong, Italy, Kuwait, Saudi Arabia, UAE, United Kingdom;

PMA numbers
P140003, P170011
Harmonized identifiers (openFDA)
Device name
Temporary Non-Roller Type Left Heart Support Blood Pump
Medical specialty description
Unknown
Device class
3
PMA number
P140003
Registration number
1220648, 2528981, 3007904765
FEI number
1220648, 2528981, 3007904765
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.