Device recalls

Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.

Abiomed, Inc. · Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circula…

Open, ClassifiedUnder Investigation by firmZ-2457-2026
Open FDA's page for this record
Cfres ID
220257
Product res number
Z-2457-2026
Event date initiated
2026-05-18
Event date posted
2026-06-18
Recall status
Open, Classified
Res event number
98946
Product code
OZD
Product description
Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.
Firm FEI number
1220648
Recalling firm
Abiomed, Inc.
Address 1
22 Cherry Hill Dr
City
Danvers
State
MA
Postal code
01923-2575
Additional info contact
Morgaine Johnson 978-646-1400
Root cause description
Under Investigation by firm
Product quantity
10 units
Distribution pattern
US distribution to GA, IA, NE, PA, TX, WA, WI. International distribution to Czech Republic, Norway, and Japan
Code info

GTIN: 00813502012279. Product Codes (Serial Number): 1000080 (677223, 672986, 673252, 644591, 645428, 644314, 613525), 1000851 (636173), 0048-0014 (499250), 0048-0024-JP (661216).

Reason for recall

Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.

Action

Johnson & Johnson MedTech/Abiomed Issued an URGETN MEDICAL DEVICE RECALL (REMOVAL) notice to its consignees on 05/18/2026 via UPS 1-day air delivery. The notice explained the reason for notification and potential patient impact Consignees were instructed to review all Impella CP Sets within inventory and quarantine affected units on hand, return all affected units using the provided return shipment label from Sedgwick, and complete and return the provided business response form. Upon receipt of returned product and completed BRF, Abiomed will provide a copy of a credit memo for the amount returned. Consignees were also instructed to forward the notification to anyone in each facility that needs to be informed and to post a copy of the notification in a visible area for awareness. If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.

PMA numbers
P140003
Harmonized identifiers (openFDA)
Device name
Temporary Non-Roller Type Left Heart Support Blood Pump
Medical specialty description
Unknown
Device class
3
PMA number
P140003
Registration number
3007904765, 2528981, 1220648
FEI number
3007904765, 2528981, 1220648
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.