Device recalls

iLet Bionic Pancreas, REF: BB1001

Beta Bionics, Inc. · The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur wh…

Open, ClassifiedSoftware designZ-2513-2026
Open FDA's page for this record
Cfres ID
218974
Product res number
Z-2513-2026
Event date initiated
2025-09-23
Event date posted
2026-06-18
Recall status
Open, Classified
Res event number
98517
Product code
QFG
Product description
iLet Bionic Pancreas, REF: BB1001
Code info
iLet ACE Pump firmware BOM000072 Rev C (version 3.3.6) and BOM000072 Rev B (version 3.3.4)
Firm FEI number
3019004087
Recalling firm
Beta Bionics, Inc.
Address 1
11 Hughes
City
Irvine
State
CA
Postal code
92618-1902
Additional info contact
Karen Hynes 855-745-3800
Root cause description
Software design
Product quantity
4580
Reason for recall

The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.

Action

On 3/12/2026, recall notices were mailed and emailed to customers who were informed of the following: On September 23, 2025, Beta Bionics, Inc. issued an Urgent Medical Device Correction notification to customers informing them of the lock screen software issue and software upgrade. These problems are resolved in newer host software versions (3.3.7 and up). Please check your software version and update your software as soon as possible. Update your software 1. Download or update your iLet app on your phone. Need help! Read the iLet App User Guide. 2. Log into your iLet app. 3. Position your iLet next to your phone with the app open. 4. Go to app settings on the iLet app and follow the on-screen instructions to update to the latest firmware. 5. Complete the acknowledgement survey at https://sprw.io/stt-3tkjI or If you have any questions, please call the firm's Customer Care at 1-855-745-3800 option 1.

Distribution pattern

US: WI, HI, OH, ME, CO, NJ, NE, KY, SC, CA, TX, NH, NY, WA, PA, MD, FL, IA, GA, MI, VT, TN, IL, IN, KS, MA, LA, MS, NC, CT, MN, VA, OK, AR, AZ, MO, MT, RI, DC, NM, OR, ID, NV, AK, ND, DE, WV, WY, SD, UT, AL

K numbers
K252770
Harmonized identifiers (openFDA)
Device name
Alternate Controller Enabled Insulin Infusion Pump
Medical specialty description
Clinical Chemistry
Regulation number
880.5730
Device class
2
K number
K213134, K234027, K263422, K261672, K231826, K251032, K253976, K250792, K253743, DEN180058, K232380, K203234, K201214, K240309, K213536, K253074, K233952, K241178, K191679, K250930, K260001, K223213, K253470, K223846, K252770, K231485, K241078, K230545, K203768, K233044
Registration number
3042027705, 3030970576, 1220477, 3013756811, 3008824097, 3016540876, 3007981285, 3005627859, 3019004087, 2029275, 2030598, 3047799708, 3016245399, 2246552, 3026266833, 3019807891, 2022435, 2032098, 3010291427, 3014585508, 2030624, 2032227, 2024024, 3014197758, 3044561902, 3016798778, 3015192982, 3032746481, 1220973
FEI number
3042027705, 3030970576, 1220477, 3013756811, 1000518731, 3008824097, 3016540876, 3000206435, 3007981285, 3005627859, 3019004087, 3047799708, 1000306431, 3016245399, 2246552, 3003058448, 3026266833, 3019807891, 2022435, 3003166194, 3010291427, 3014585508, 2024024, 3014197758, 3044561902, 3016798778, 3015192982, 3032746481, 1220973
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.