Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A P…
Inspire Medical Systems Inc. · Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a prod…
Open FDA's page for this record- Cfres ID
- 220369
- Product res number
- Z-2515-2026
- Event date initiated
- 2026-05-11
- Event date posted
- 2026-06-18
- Recall status
- Open, Classified
- Res event number
- 98990
- Product code
- MNQ
- Firm FEI number
- 3007666314
- Recalling firm
- Inspire Medical Systems Inc.
- Address 1
- 5500 Wayzata Blvd Ste 1600 5500 Wayzata Blvd
- Address 2
- Ste 1600
- City
- Golden Valley
- State
- MN
- Postal code
- 55416-1237
- Additional info contact
- 763-205-7974
- Root cause description
- Process control
- Product quantity
- 50 impacted units
- Distribution pattern
- Worldwide distribution - US Nationwide and the country of Hong Kong.
- Product description
Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A
- Code info
Lot Code: Model No 4340 (USA only lead); UDI-DI [4340] 10855728005946; Inspire part number [4340] 900-014-003; serial numbers T18806, T45012, T45017, T49002, T67098, T67152, T69541, T69542, T69553, T69558, T69562 Model No 4340 (Global lead); UDI-DI [4340] 10810098650301; Inspire part number [4340] 900-014-005; serial numbers T91485, T91486, T91491, T91495, T91513, T91520, T91521, T91527, T91533, T91537, T91540, T91545, T96165, T96168, T96192, T96193, T96195, T96225, T96237, T98118, T98126, T98136, T98137, T98138, T98141, T98143, T98145, T98150, T98153, T98156, T98163, T98165, T98176, T98177, T98178, T98181, T98183, T98190, T96217
- Reason for recall
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
- Action
Firm began notifying consignees on May 11, 2026. Letters were titled "URGENT Medical Device Correction." The firm requests consignees' assistance in removing affected product from clinical inventory and returning them to Inspire.
- PMA numbers
- P130008
- Harmonized identifiers (openFDA)
- Device name
- Stimulator, Hypoglossal Nerve, Implanted, Apnea
- Medical specialty description
- Unknown
- Device class
- 3
- PMA number
- P250013, P130008, P240024
- Registration number
- 3008851924, 3030499299, 3003857828, 3005672861, 3009183442, 2183613, 2183787, 2183744, 3030384402, 3024996452, 3010376846, 3017191859, 3007666314, 1047843, 3030384719, 3036471728
- FEI number
- 3008851924, 3030499299, 3003857828, 3005672861, 3009183442, 2183613, 2183787, 2183744, 3030384402, 3024996452, 3010376846, 3017191859, 3007666314, 1047843, 3030384719, 3036471728