Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspir…
Inspire Medical Systems Inc. · Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a prod…
Open FDA's page for this record- Cfres ID
- 220607
- Product res number
- Z-2516-2026
- Event date initiated
- 2026-05-11
- Event date posted
- 2026-06-18
- Recall status
- Open, Classified
- Res event number
- 98990
- Product code
- MNQ
- Firm FEI number
- 3007666314
- Recalling firm
- Inspire Medical Systems Inc.
- Address 1
- 5500 Wayzata Blvd Ste 1600 5500 Wayzata Blvd
- Address 2
- Ste 1600
- City
- Golden Valley
- State
- MN
- Postal code
- 55416-1237
- Additional info contact
- 763-205-7974
- Root cause description
- Process control
- Product quantity
- 10 impacted units
- Distribution pattern
- Worldwide distribution - US Nationwide and the country of Hong Kong.
- Product description
Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire Stimulation Lead (Model 4063) is designed to deliver stimulation to the hypoglossal nerve for the treatment of obstructive sleep apnea. The lead features a flexible, self-sizing cuff. Electrodes in the inner surface of the cuff deliver stimulation to the nerve. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A
- Code info
Lot Code: Model 4063 (USA only lead); UDI-DI [4063] 10855728005793; Inspire part number [4063] 900-013-001; serial numbers D53393, D61104 Model 4063 (Global lead); UDI-DI [4063] 10810098650295, Inspire part number [4063] 900-013-003; serial numbers D89539, D89540, D89543, D89559, D89563, D89564, D89577, D8959
- Reason for recall
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
- Action
Firm began notifying consignees on May 11, 2026. Letters were titled "URGENT Medical Device Correction." The firm requests consignees' assistance in removing affected product from clinical inventory and returning them to Inspire.
- PMA numbers
- P130008
- Harmonized identifiers (openFDA)
- Device name
- Stimulator, Hypoglossal Nerve, Implanted, Apnea
- Medical specialty description
- Unknown
- Device class
- 3
- PMA number
- P250013, P130008, P240024
- Registration number
- 3008851924, 3030499299, 3003857828, 3005672861, 3009183442, 2183613, 2183787, 2183744, 3030384402, 3024996452, 3010376846, 3017191859, 3007666314, 1047843, 3030384719, 3036471728
- FEI number
- 3008851924, 3030499299, 3003857828, 3005672861, 3009183442, 2183613, 2183787, 2183744, 3030384402, 3024996452, 3010376846, 3017191859, 3007666314, 1047843, 3030384719, 3036471728