Device recalls

Splashwire Hydrophilic Guide Wire: REF: MSWSTDA35150 UDI-DI codes: 00884450529645 00884450529669 00884450529690 REF: MSWSTDA35180 UDI-DI co…

Merit Medical Systems, Inc. · Due to unsealed portions of pouches.

Open, ClassifiedProcess designZ-2517-2026
Open FDA's page for this record
Cfres ID
219884
Product res number
Z-2517-2026
Event date initiated
2026-05-08
Event date posted
2026-06-18
Recall status
Open, Classified
Res event number
98818
Product code
DQX
Firm FEI number
1721504
Recalling firm
Merit Medical Systems, Inc.
Address 1
1600 W Merit Pkwy
City
South Jordan
State
UT
Postal code
84095-2416
Additional info contact
Mr. Adrian Cole 801-253-1600
Reason for recall
Due to unsealed portions of pouches.
Root cause description
Process design
Product quantity
5,004
Product description

Splashwire Hydrophilic Guide Wire: REF: MSWSTDA35150 UDI-DI codes: 00884450529645 00884450529669 00884450529690 REF: MSWSTDA35180 UDI-DI codes: 00884450529645 00884450529669 REF: MSWSTDA35260EX UDI-DI code: 00884450529683 REF: MSWSTDA3580 UDI-DI codes: 00884450529645 00884450529690 00884450529744 REF: MSWSTDS35150 UDI-DI codes: 00884450529690 00884450529744 REF: MSWSTFA35260EX UDI-DI code: 00884450529447 REF: MSWSTFA3580 UDI-DI code: 00884450529515 REF: MSWSTFS35150 UDI-DI code: 00884450529553 The Merit Hydrophilic Guide Wire is intended to be used in the peripheral vascular system to facilitate the placement of devices during diagnostic and interventional procedures.

Code info

REF: MSWSTDA35150 UDI-DI code: 00884450529645 Lot Numbers: K3410951 K3420005 K3420006 K3420007 K3420106 UDI-DI code: 00884450529669 Lot Number: K3420005 UDI-DI code: 00884450529690 Lot Number: K3420006 REF: MSWSTDA35180 UDI-DI code: 00884450529645 Lot Number: K3405121 UDI-DI code: 00884450529669 Lot Numbers: K3405121 K3405126 K3405127 K3405128 K3405129 K3405130 K3410953 K3420010 REF: MSWSTDA35260EX UDI-DI code: 00884450529683 Lot Numbers: K3405122 K3410937 K3410958 REF: MSWSTDA3580 UDI-DI code: 00884450529645 Lot Number: K3405257 UDI-DI code: 00884450529690 Lot Number: K3405257 UDI-DI code: 00884450529744 Lot Number: K3405257 REF: MSWSTDS35150 UDI-DI code: 00884450529690 Lot Number: K3420082 UDI-DI code: 00884450529744 Lot Numbers: K3420082 K3420101 REF: MSWSTFA35260EX UDI-DI code: 00884450529447 Lot Number: K3405261 REF: MSWSTFA3580 UDI-DI code: 00884450529515 Lot Number: K3420079 REF: MSWSTFS35150 UDI-DI code: 00884450529553 Lot Number: K3399729

Action

On 05/08/2026, the firm sent via FedEx Express an "U.S. URGENT MEDICAL DEVICE RECALL NOTICE" to inform customers that internal review identified that certain devices manufactured during a defined time period may have an incomplete seal in the pouch of the Merit SplashWire Hydrophilic Steerable Guide Wire. Customer are instructed to: 1. Immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within their facility. 2. A picture of the SplashWire Hydrophilic Steerable Guide Wire Hydrophilic Steerable Guide Wire is attached to help them identify the product. 3. Immediately discontinue the use of the affected lots of the SplashWire Hydrophilic Steerable Guide Wire. 4. Ensure that applicable personnel within their organization are made aware of this recall. 5. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 6. Fill out, scan and email the completed Customer Response Form to Customer Service at RESPONSE@merit.com within 10 business days. All affected product shipped to them must be accounted for on the CRF. 7. Immediately return all affected lots in their possession to Merit, per the instructions in the attached CRF. 8. When requesting replacement for the SplashWire Hydrophilic Steerable Guide Wire please utilize the table for the replacement part number and contact RESPONSE@merit.com, when they provide their completed Customer Response Form. For questions - contact your Merit Sales Representative or Merit Customer Service via email at RESPONSE@merit.com or via phone at +1 800 356 3748 | Hours: 6 am to 6 pm MST | Mon-Fri.

Distribution pattern

Worldwide - U.S. Nationwide distribution in the states and territory AL, AZ, CA, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and Puerto Rico. The countries of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece, Guatemala, Ireland, Italy, Kazakhstan, Kenya, Netherlands, Norway, Portugal, Spain, Sweden, and Switzerland.

K numbers
K201595
Harmonized identifiers (openFDA)
Device name
Wire, Guide, Catheter
Medical specialty description
Cardiovascular
Regulation number
870.1330
Device class
2
K number
K862983, K080637, K050964, K113454, K032899, K130766, K912293, K140673, K083904, K852123, K081021, K093647, K192454, K261640, K124007, K040100, K943522, K021323, K970244, K100710, K021638, K955135, K871492, K874542, K903923, K052022, K902153, K042338, K863178, K021228, K102818, K171897, K031277, K884274, K850828, K011905, K991646, K861574, K030617, K173277 +729 more
Registration number
9616088, 1016427, 3020968442, 3014579161, 3005941719, 3009380063, 1000121056, 3013611763, 3002808486, 3036802499, 3014279513, 3007830707, 2134812, 3010432890, 3012223769, 3006950086, 3006283319, 3011564431, 3015489752, 3011137372, 3014590708, 3002917664, 3007138831, 3012536737, 3012931345, 3007635982, 1721686, 3012159165, 1417485, 3003084417, 3013666336, 2024168, 3000126629, 1319639, 3004748528, 1644312, 9610849, 3012154226, 8010026, 2183870 +221 more
FEI number
3003737899, 3002808331, 1016427, 3020968442, 3014579161, 3009380063, 1000121056, 3013611763, 3002808486, 3036802499, 3014279513, 3002902735, 3007830707, 3010432890, 3006950086, 1000222374, 3011564431, 3015489752, 3014590708, 3007138831, 3002917664, 3012536737, 3012931345, 3004635528, 3007635982, 1721686, 3001452571, 3012159165, 1000139754, 1417485, 3003084417, 3013666336, 2024168, 3000126629, 3002807095, 3004748528, 3002806565, 3012154226, 2183870, 3035642068 +220 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.