Philips Avalon Fetal Monitor, FM 30 Part numberM2703A
Philips North America Llc · Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or…
Open FDA's page for this record- Cfres ID
- 220770
- Product res number
- Z-2523-2026
- Event date initiated
- 2026-05-26
- Event date posted
- 2026-06-22
- Recall status
- Open, Classified
- Res event number
- 99084
- Product code
- HGM
- Product description
- Philips Avalon Fetal Monitor, FM 30 Part numberM2703A
- Firm FEI number
- 3016618143
- Recalling firm
- Philips North America Llc
- Address 1
- 222 Jacobs St
- City
- Cambridge
- State
- MA
- Postal code
- 02141-2289
- Additional info contact
- Philips Customer Services 800-722-9377
- Root cause description
- Process control
- Product quantity
- 169 units
- Code info
UDI/DI 00884838000414, Serial Numbers: DE79646555, DE79646616, DE79646645, DE79646647, DE79646651, DE79646697, DE79646702, DE79646746, DE79646754, DE79646757, DE79646758, DE79646762, DE79646765, DE79646775, DE79646776, DE79646783, DE79646857, DE79646891, DE79646924, DE79646750, DE79646751, DE79646756, DE79646804, DE79646592, DE79646877, DE79646644, DE79646706, DE79646723, DE79646653, DE79646766, DE79646786, DE79646824, DE79646827, DE79646848, DE79646956, DE79646709, DE79646892, DE79646724, DE79646729, DE79646732, DE79646733, DE79646927, DE79646675, DE79646820, DE79646821, DE79646837, DE79646866, DE79646869, DE79646871, DE79646875, DE79646882, DE79646683, DE79646676, DE79646640, DE79646849, DE79646952, DE79646699, DE79646694, DE79646648, DE79646701, DE79646900, DE79647027, DE79647039, DE79646654, DE79646800, DE79646604, DE79646660, DE79646829, DE79646747, DE79646606, DE79646826, DE79646784, DE79646794.
- Reason for recall
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
- Action
Philips issued an URGENT Medical Device Correction notice to its consignees on 5/26/2026 via USPS certified mail. The notice explained the issue and how it can be identified, potential risk to the patient should the loss of audio is not detected, and requested the following: "Actions that should be taken by the customer / user in order to prevent risks for patients or users " Ensure fetal heart rate tones are enabled on the Avalon Fetal Monitor until the planned fix/correction is implemented. " Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. " Place this Urgent Medical Device Correction with the documentation of the Philips Avalon Fetal Monitor and associated devices in a place where it is most likely to be seen and viewed."
- Distribution pattern
Worldwide distribution - US Nationwide and The countries of AE, AT, AU, BE, BG, BH, BR, CA, CH, CN, CR, CZ, DE, ES, FI, FR, GB, GR, HK, ID, IL, IN, IT, JO, JP, KR, KR, LT, LU, LV, MX, MY, NL, NZ, OM, SA, SE, SG, SI, TR, TW, ZA.
- K numbers
- K140535
- Harmonized identifiers (openFDA)
- Device name
- System, Monitoring, Perinatal
- Medical specialty description
- Obstetrics/Gynecology
- Regulation number
- 884.2740
- Device class
- 2
- K number
- K935301, K991739, K040903, K991905, K080226, K052715, K771195, K892501, K233440, K200975, K983395, K032252, K050234, K012718, K812863, K831113, K992607, K932077, K011304, K032772, K781359, K960109, K852704, K894958, K051175, K103172, K173042, K132650, K791035, K832849, K821580, K895045, K852608, K001677, K002150, K761201, K851252, K073221, K823188, K843918 +175 more
- Registration number
- 8020045, 3007734888, 3010611950, 3003971136, 3013500228, 3010390468, 3036470679, 3016618143, 1000589001, 1930027, 2028916, 3013400897, 3010698831, 3012359877, 3009499478, 3009077524, 3021635975, 3001570235, 3015529350, 3007309299, 3014541700, 3003681187, 9680215, 3000126629, 3004423584, 1450908, 9610105, 9710602, 3013596024, 3013783488, 9681377, 3007048259, 3016701404, 9615396, 3015976667, 1218950, 3008729547, 3014220748, 3006179052, 3007607999 +16 more
- FEI number
- 3007734888, 3010611950, 3003971136, 3013500228, 3010390468, 3036470679, 3016618143, 3003513129, 1000589001, 1930027, 1000300598, 3013400897, 3010698831, 3003190524, 3012359877, 1000142646, 3009499478, 3009077524, 3021635975, 3001570235, 3003015552, 3002687025, 3015529350, 3007309299, 3014541700, 3003681187, 3002807167, 3000126629, 3004423584, 1450908, 3013596024, 3001084824, 3013783488, 1000140026, 3007048259, 3016701404, 3015976667, 1218950, 3008729547, 3014220748 +16 more