Device recalls

Philips Avalon Fetal Monitor, FM 30 Part numberM2703A

Philips North America Llc · Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or…

Open, ClassifiedProcess controlZ-2523-2026
Open FDA's page for this record
Cfres ID
220770
Product res number
Z-2523-2026
Event date initiated
2026-05-26
Event date posted
2026-06-22
Recall status
Open, Classified
Res event number
99084
Product code
HGM
Product description
Philips Avalon Fetal Monitor, FM 30 Part numberM2703A
Firm FEI number
3016618143
Recalling firm
Philips North America Llc
Address 1
222 Jacobs St
City
Cambridge
State
MA
Postal code
02141-2289
Additional info contact
Philips Customer Services 800-722-9377
Root cause description
Process control
Product quantity
169 units
Code info

UDI/DI 00884838000414, Serial Numbers: DE79646555, DE79646616, DE79646645, DE79646647, DE79646651, DE79646697, DE79646702, DE79646746, DE79646754, DE79646757, DE79646758, DE79646762, DE79646765, DE79646775, DE79646776, DE79646783, DE79646857, DE79646891, DE79646924, DE79646750, DE79646751, DE79646756, DE79646804, DE79646592, DE79646877, DE79646644, DE79646706, DE79646723, DE79646653, DE79646766, DE79646786, DE79646824, DE79646827, DE79646848, DE79646956, DE79646709, DE79646892, DE79646724, DE79646729, DE79646732, DE79646733, DE79646927, DE79646675, DE79646820, DE79646821, DE79646837, DE79646866, DE79646869, DE79646871, DE79646875, DE79646882, DE79646683, DE79646676, DE79646640, DE79646849, DE79646952, DE79646699, DE79646694, DE79646648, DE79646701, DE79646900, DE79647027, DE79647039, DE79646654, DE79646800, DE79646604, DE79646660, DE79646829, DE79646747, DE79646606, DE79646826, DE79646784, DE79646794.

Reason for recall

Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.

Action

Philips issued an URGENT Medical Device Correction notice to its consignees on 5/26/2026 via USPS certified mail. The notice explained the issue and how it can be identified, potential risk to the patient should the loss of audio is not detected, and requested the following: "Actions that should be taken by the customer / user in order to prevent risks for patients or users " Ensure fetal heart rate tones are enabled on the Avalon Fetal Monitor until the planned fix/correction is implemented. " Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. " Place this Urgent Medical Device Correction with the documentation of the Philips Avalon Fetal Monitor and associated devices in a place where it is most likely to be seen and viewed."

Distribution pattern

Worldwide distribution - US Nationwide and The countries of AE, AT, AU, BE, BG, BH, BR, CA, CH, CN, CR, CZ, DE, ES, FI, FR, GB, GR, HK, ID, IL, IN, IT, JO, JP, KR, KR, LT, LU, LV, MX, MY, NL, NZ, OM, SA, SE, SG, SI, TR, TW, ZA.

K numbers
K140535
Harmonized identifiers (openFDA)
Device name
System, Monitoring, Perinatal
Medical specialty description
Obstetrics/Gynecology
Regulation number
884.2740
Device class
2
K number
K935301, K991739, K040903, K991905, K080226, K052715, K771195, K892501, K233440, K200975, K983395, K032252, K050234, K012718, K812863, K831113, K992607, K932077, K011304, K032772, K781359, K960109, K852704, K894958, K051175, K103172, K173042, K132650, K791035, K832849, K821580, K895045, K852608, K001677, K002150, K761201, K851252, K073221, K823188, K843918 +175 more
Registration number
8020045, 3007734888, 3010611950, 3003971136, 3013500228, 3010390468, 3036470679, 3016618143, 1000589001, 1930027, 2028916, 3013400897, 3010698831, 3012359877, 3009499478, 3009077524, 3021635975, 3001570235, 3015529350, 3007309299, 3014541700, 3003681187, 9680215, 3000126629, 3004423584, 1450908, 9610105, 9710602, 3013596024, 3013783488, 9681377, 3007048259, 3016701404, 9615396, 3015976667, 1218950, 3008729547, 3014220748, 3006179052, 3007607999 +16 more
FEI number
3007734888, 3010611950, 3003971136, 3013500228, 3010390468, 3036470679, 3016618143, 3003513129, 1000589001, 1930027, 1000300598, 3013400897, 3010698831, 3003190524, 3012359877, 1000142646, 3009499478, 3009077524, 3021635975, 3001570235, 3003015552, 3002687025, 3015529350, 3007309299, 3014541700, 3003681187, 3002807167, 3000126629, 3004423584, 1450908, 3013596024, 3001084824, 3013783488, 1000140026, 3007048259, 3016701404, 3015976667, 1218950, 3008729547, 3014220748 +16 more