VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001

Ventec Life Systems, Inc. · Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which m…

Open, ClassifiedVendor change controlZ-2525-2026
Open FDA's page for this record
Cfres ID
220084
Product res number
Z-2525-2026
Event date initiated
2026-05-08
Event date posted
2026-06-23
Recall status
Open, Classified
Res event number
98888
Product code
CBK
Product description
VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
Code info
REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613
Firm FEI number
3013095415
Recalling firm
Ventec Life Systems, Inc.
Address 1
22002 26th Ave Se
City
Bothell
State
WA
Postal code
98021-4903
Additional info contact
425-686-1722
Root cause description
Vendor change control
Product quantity
1
Distribution pattern
US Nationwide distribution in the state of UT.
Reason for recall

Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.

Action

On 5/8/2026, recall notices were mailed to customers who were asked to do the following: 1) Immediately discontinue use of affected devices. 2) Inspect current stock, quarantine/segregate any unused affected products, remove affected devices from clinical service and segregate to prevent further use. 3) Use an unaffected ventilator for patient support. 4) If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification. 5) Confirm receipt of this notification letter by returning the enclosed Customer Response Form via email to Ventec-Recall@reacthealth.com Upon receipt of the Customer Response Form, firm will arrange for repair of existing device(s). Return the affected product to React Health If you have any questions, please contact firm at 844-698-6276.

K numbers
K162877
Harmonized identifiers (openFDA)
Device name
Ventilator, Continuous, Facility Use
Medical specialty description
Anesthesiology
Regulation number
868.5895
Device class
2
K number
K093886, K022674, K923444, K083693, K922314, K192307, K950484, K914607, K101850, K123637, K062793, K880923, K960964, K905235, K060705, K800114, K895513, K924930, K012177, K081670, K253647, K081521, K102120, K090468, K781160, K844350, K911804, K001646, K233486, K772065, K180098, K113743, K201306, K252663, K092148, K042468, K131774, K201874, K924043, K860798 +465 more
Registration number
3016723884, 3010317211, 2024500, 3009302812, 3009609218, 2025816, 3013154126, 3029845, 8021084, 3006783791, 2221819, 3038259592, 3049808762, 3004428458, 2518422, 9680215, 3000126629, 1922553, 9610849, 3007344957, 3029922922, 3008867836, 3001421318, 3026740087, 3007603826, 3007573469, 3010157426, 3031369651, 3010817335, 3005783425, 3003135857, 9615827, 3009107628, 3029871637, 3018094310, 3014631252, 2085602, 9617566, 2518436, 1225492 +91 more
FEI number
3003244771, 2995, 3016723884, 3010317211, 2024500, 3003513129, 3009302812, 3009609218, 2025816, 3000210616, 3013154126, 3006783791, 2221819, 3038259592, 3049808762, 3004428458, 2518422, 3000126629, 1922553, 3007344957, 3029922922, 3008867836, 3001084824, 3001421318, 3004213967, 3026740087, 3007603826, 3010157426, 3000957797, 3031369651, 3010817335, 3005783425, 3003135857, 3002807626, 3009107628, 3029871637, 3018094310, 3014631252, 2518436, 1000524541 +91 more