VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
Ventec Life Systems, Inc. · Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which m…
Open FDA's page for this record- Cfres ID
- 220084
- Product res number
- Z-2525-2026
- Event date initiated
- 2026-05-08
- Event date posted
- 2026-06-23
- Recall status
- Open, Classified
- Res event number
- 98888
- Product code
- CBK
- Product description
- VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
- Code info
- REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613
- Firm FEI number
- 3013095415
- Recalling firm
- Ventec Life Systems, Inc.
- Address 1
- 22002 26th Ave Se
- City
- Bothell
- State
- WA
- Postal code
- 98021-4903
- Additional info contact
- 425-686-1722
- Root cause description
- Vendor change control
- Product quantity
- 1
- Distribution pattern
- US Nationwide distribution in the state of UT.
- Reason for recall
Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
- Action
On 5/8/2026, recall notices were mailed to customers who were asked to do the following: 1) Immediately discontinue use of affected devices. 2) Inspect current stock, quarantine/segregate any unused affected products, remove affected devices from clinical service and segregate to prevent further use. 3) Use an unaffected ventilator for patient support. 4) If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification. 5) Confirm receipt of this notification letter by returning the enclosed Customer Response Form via email to Ventec-Recall@reacthealth.com Upon receipt of the Customer Response Form, firm will arrange for repair of existing device(s). Return the affected product to React Health If you have any questions, please contact firm at 844-698-6276.
- K numbers
- K162877
- Harmonized identifiers (openFDA)
- Device name
- Ventilator, Continuous, Facility Use
- Medical specialty description
- Anesthesiology
- Regulation number
- 868.5895
- Device class
- 2
- K number
- K093886, K022674, K923444, K083693, K922314, K192307, K950484, K914607, K101850, K123637, K062793, K880923, K960964, K905235, K060705, K800114, K895513, K924930, K012177, K081670, K253647, K081521, K102120, K090468, K781160, K844350, K911804, K001646, K233486, K772065, K180098, K113743, K201306, K252663, K092148, K042468, K131774, K201874, K924043, K860798 +465 more
- Registration number
- 3016723884, 3010317211, 2024500, 3009302812, 3009609218, 2025816, 3013154126, 3029845, 8021084, 3006783791, 2221819, 3038259592, 3049808762, 3004428458, 2518422, 9680215, 3000126629, 1922553, 9610849, 3007344957, 3029922922, 3008867836, 3001421318, 3026740087, 3007603826, 3007573469, 3010157426, 3031369651, 3010817335, 3005783425, 3003135857, 9615827, 3009107628, 3029871637, 3018094310, 3014631252, 2085602, 9617566, 2518436, 1225492 +91 more
- FEI number
- 3003244771, 2995, 3016723884, 3010317211, 2024500, 3003513129, 3009302812, 3009609218, 2025816, 3000210616, 3013154126, 3006783791, 2221819, 3038259592, 3049808762, 3004428458, 2518422, 3000126629, 1922553, 3007344957, 3029922922, 3008867836, 3001084824, 3001421318, 3004213967, 3026740087, 3007603826, 3010157426, 3000957797, 3031369651, 3010817335, 3005783425, 3003135857, 3002807626, 3009107628, 3029871637, 3018094310, 3014631252, 2518436, 1000524541 +91 more