Device recalls

Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310

PFM MEDICAL INC. · A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requi…

Open, ClassifiedUnder Investigation by firmZ-2526-2026
Open FDA's page for this record
Cfres ID
220251
Product res number
Z-2526-2026
Event date initiated
2026-04-21
Event date posted
2026-06-23
Recall status
Open, Classified
Res event number
98942
Product code
DXE
Product description
Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310
Code info
REF/UDI-DI/Lot(Expiration): 147305/04042301061829/1049868(10/14/2028); 147310/04042301033567/1049960(10/20/28)
Firm FEI number
3013666218
Recalling firm
PFM MEDICAL INC.
Address 1
1916 Palomar Oaks Way
City
Carlsbad
State
CA
Postal code
92008-5523
Additional info contact
760-758-8749 Ext. 222
Root cause description
Under Investigation by firm
Product quantity
32
Distribution pattern
US Nationwide distribution in the states of NY, MD.
Reason for recall

A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.

Action

On 4/21/2026, recall notices were emailed to customers who were asked to do the following: 1. Inform all customers (hospitals and clinics) and device users (physicians) who have received or used affected batches. This notification should also be shared with any organizations where the potentially affected devices have been transferred. 2. Return unused devices from the affected batches to PFM Medical, Inc. 3. Complete the enclosed Recall Response Form and email back to recall@pfmmedicalusa.com

K numbers
K011783
Harmonized identifiers (openFDA)
Device name
Catheter, Embolectomy
Medical specialty description
Cardiovascular
Regulation number
870.5150
Device class
2
K number
K982141, K241330, K894681, K233820, K233819, K935405, K011783, K974335, K905139, K202049, K961883, K883852, K910372, K081573, K832023, K973465, K992933, K972511, K864062, K820646, K872485, K163031, K892410, K864007, K770714, K240078, K864330, K002627, K955125, K152335, K213771, K761031, K971314, K891021, K850015, K020634, K920823, K973477, K022145, K902327 +34 more
Registration number
3015309643, 3010041511, 3035111606, 3009051888, 8043983, 3006082230, 2015691, 3032916632, 1820334, 3012050423, 3015225571, 1220948, 3030714660, 3040153768, 3045042524, 3006950086, 3011137372, 3015859709, 1061124, 3012536737, 3005704822, 1721686, 1000393132, 2522007, 1319639, 2648045, 2183870, 3033589330, 2027111, 9616662, 2029386, 1721504, 9680794, 2183744, 1724474, 3008496528, 3013666218, 1018233, 2024311, 9615817 +2 more
FEI number
1000138054, 3015309643, 3010041511, 3003081151, 3035111606, 3003737899, 3009051888, 3006082230, 2015691, 3032916632, 1820334, 3012050423, 3015225571, 3000335778, 3003769549, 3030714660, 1220948, 3040153768, 3045042524, 3006950086, 1000512168, 3015859709, 3012536737, 3005704822, 1721686, 3003094851, 1000393132, 2522007, 1220452, 2648045, 2183870, 3033589330, 2027111, 3005747797, 1721504, 2183744, 3013666218, 3008496528, 1018233, 1000306647 +2 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.