Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310
PFM MEDICAL INC. · A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requi…
Open FDA's page for this record- Cfres ID
- 220251
- Product res number
- Z-2526-2026
- Event date initiated
- 2026-04-21
- Event date posted
- 2026-06-23
- Recall status
- Open, Classified
- Res event number
- 98942
- Product code
- DXE
- Product description
- Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310
- Code info
- REF/UDI-DI/Lot(Expiration): 147305/04042301061829/1049868(10/14/2028); 147310/04042301033567/1049960(10/20/28)
- Firm FEI number
- 3013666218
- Recalling firm
- PFM MEDICAL INC.
- Address 1
- 1916 Palomar Oaks Way
- City
- Carlsbad
- State
- CA
- Postal code
- 92008-5523
- Additional info contact
- 760-758-8749 Ext. 222
- Root cause description
- Under Investigation by firm
- Product quantity
- 32
- Distribution pattern
- US Nationwide distribution in the states of NY, MD.
- Reason for recall
A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
- Action
On 4/21/2026, recall notices were emailed to customers who were asked to do the following: 1. Inform all customers (hospitals and clinics) and device users (physicians) who have received or used affected batches. This notification should also be shared with any organizations where the potentially affected devices have been transferred. 2. Return unused devices from the affected batches to PFM Medical, Inc. 3. Complete the enclosed Recall Response Form and email back to recall@pfmmedicalusa.com
- K numbers
- K011783
- Harmonized identifiers (openFDA)
- Device name
- Catheter, Embolectomy
- Medical specialty description
- Cardiovascular
- Regulation number
- 870.5150
- Device class
- 2
- K number
- K982141, K241330, K894681, K233820, K233819, K935405, K011783, K974335, K905139, K202049, K961883, K883852, K910372, K081573, K832023, K973465, K992933, K972511, K864062, K820646, K872485, K163031, K892410, K864007, K770714, K240078, K864330, K002627, K955125, K152335, K213771, K761031, K971314, K891021, K850015, K020634, K920823, K973477, K022145, K902327 +34 more
- Registration number
- 3015309643, 3010041511, 3035111606, 3009051888, 8043983, 3006082230, 2015691, 3032916632, 1820334, 3012050423, 3015225571, 1220948, 3030714660, 3040153768, 3045042524, 3006950086, 3011137372, 3015859709, 1061124, 3012536737, 3005704822, 1721686, 1000393132, 2522007, 1319639, 2648045, 2183870, 3033589330, 2027111, 9616662, 2029386, 1721504, 9680794, 2183744, 1724474, 3008496528, 3013666218, 1018233, 2024311, 9615817 +2 more
- FEI number
- 1000138054, 3015309643, 3010041511, 3003081151, 3035111606, 3003737899, 3009051888, 3006082230, 2015691, 3032916632, 1820334, 3012050423, 3015225571, 3000335778, 3003769549, 3030714660, 1220948, 3040153768, 3045042524, 3006950086, 1000512168, 3015859709, 3012536737, 3005704822, 1721686, 3003094851, 1000393132, 2522007, 1220452, 2648045, 2183870, 3033589330, 2027111, 3005747797, 1721504, 2183744, 3013666218, 3008496528, 1018233, 1000306647 +2 more