Device recalls

Off-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444

Zimmer, Inc. · Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.

Open, ClassifiedUnder Investigation by firmZ-2530-2026
Open FDA's page for this record
Cfres ID
220692
Product res number
Z-2530-2026
Event date initiated
2026-05-20
Event date posted
2026-06-23
Recall status
Open, Classified
Res event number
99045
Product code
PHX
Product description
Off-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444
Code info
Lot Code: 110040444 ; UDI-DI (01)00889024681903(17)350531(10)097487 ; Lot Number 097487
Firm FEI number
1000220733
Recalling firm
Zimmer, Inc.
Address 1
1800 W Center St
City
Warsaw
State
IN
Postal code
46580-2304
Additional info contact
411 Technical Services 800-3482759
Reason for recall
Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
Root cause description
Under Investigation by firm
Product quantity
11 units
Distribution pattern
US Nationwide distribution.
Action

An URGENT MEDICAL DEVICE RECALL CORRECTION notification letter dated 5/26/26 was sent to customers. Distributor Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement for each return and send to CorporateQuality.PostMarket@zimmerbiomet.com. 3. Return the Additional Accounts form to CorporateQuality.PostMarket@zimmerbiomet.com. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons who are current users of the affected products that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. 4. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-348-2759 between 8:00 a.m. and 5:00 p.m. Eastern, Monday through Friday. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Risk Manager Responsibilities: 1. Review this correction notice and ensure that affected personnel are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 3. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 4. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-348-2759 between 8:00 am and 5:00pm EST, Monday through Friday. Alternatively, y

K numbers
K241043
Harmonized identifiers (openFDA)
Device name
Shoulder Prosthesis, Reverse Configuration
Medical specialty description
Orthopedic
Regulation number
888.3660
Device class
2
K number
K121183, K180632, K261016, K202098, K234105, K050316, K212415, K053274, K254003, K210533, K150568, K113254, K152754, K063569, K182536, K193394, K251184, K130550, K214001, K111061, K232583, K133378, K260556, K100741, K052906, K182039, K093803, K161120, K211074, K082702, K223833, K213117, K183077, K102951, K173900, K110708, K193373, K253624, K250644, K243643 +167 more
Registration number
1038671, 1649518, 3003998208, 2249615, 3002579136, 3013791180, 3006721341, 3011500362, 3009546990, 3014302784, 3013552417, 1644408, 9616671, 1226544, 3013014058, 3033273230, 2245304, 1043534, 3004641308, 3013561205, 3033509898, 3009555440, 3002806470, 3025603301, 3004748528, 3010400367, 3005497913, 3014833750, 3010331645, 3007441485, 1836357, 1721676, 3005061536, 1526534, 3010097171, 3006128100, 3009185591, 1422572, 3003403260, 3006017180 +153 more
FEI number
1038671, 3002808293, 3003998208, 3002579136, 3013791180, 3006721341, 3011500362, 1000306303, 3009546990, 3014302784, 3000215273, 3013552417, 3003064967, 3013014058, 3033273230, 2245304, 1043534, 3004641308, 3013561205, 3033509898, 3009555440, 3002806470, 3025603301, 3004748528, 3010400367, 3005497913, 3003070421, 3014833750, 3001452082, 3010331645, 3007441485, 1721676, 3003414678, 3005061536, 1526534, 3010097171, 3006128100, 3009185591, 3003403260, 1000149660 +153 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.