INZONE DETACHMENT SYSTEM, REF: M00345100950
Stryker Neurovascular · Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visu…
Open FDA's page for this record- Cfres ID
- 219266
- Product res number
- Z-2552-2026
- Event date initiated
- 2026-04-06
- Event date posted
- 2026-06-24
- Recall status
- Open, Classified
- Res event number
- 98630
- Product code
- HCG
- Product description
- INZONE DETACHMENT SYSTEM, REF: M00345100950
- Firm FEI number
- 3008853977
- Recalling firm
- Stryker Neurovascular
- Address 1
- 47900 Bayside Pkwy
- City
- Fremont
- State
- CA
- Postal code
- 94538-6515
- Additional info contact
- 415-872-6400
- Root cause description
- Component design/selection
- Product quantity
- 44,937
- Code info
UDI-DI: 04546540697950. Lots: WMP133638, WMP133871, WMP133872, WMP133988, WMP133989, WMP134016, WMP134132, WMP134133, WMP134234, WMP134235, WMP134255, WMP134256, WMP134380, WMP134381, WMP134432, WMP134433, WMP134518, WMP134519, WMP134609, WMP134610, WMP134794, WMP134795, WMP134939, WMP134940, WMP135187, WMP135243, WMP135302, WMP135314, WMP135354, WMP135355, WMP135387, WMP135417, WMP135473, WMP135475, WMP135491, WMP135492, WMP135623, WMP135720, WMP135727, WMP135807, WMP135865, WMP135897, WMP136084, WMP136107, WMP136190, WMP136200, WMP136611, WMP136642, WMP136736, WMP136748, WMP136852, WMP136894, WMP136985
- Reason for recall
Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.
- Action
On 4/3/2026, recall notices were mailed to Supply Chain Management and Recall Coordinators who were asked to do the following: 1) Segregate the affected devices in a secure location for return to recalling firm. 2) If any of the subject devices have been distributed to other organizations, please forward a copy of this notice to the new users and provide contact details to recalling firm so new recipients can be informed appropriately. 3) Maintain awareness of this communication internally until all required actions have been completed within your facility. 4) Complete and return the Business Reply Form via email to nvfieldactions@stryker.com For any additional information or questions, please contact nvfieldactions@stryker.com or US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: NVCustomerCare@stryker.com
- Distribution pattern
Worldwide - US Nationwide distribution including in the states of PA, AZ, FL, CA, IL, AK, NY, MI, NJ, WI, SD, KY, AR, TX, AL, TN, MO, MA, IN, NE, WV, MT, CT, NC, VA, NV, CO, DC, DE, OH, ME, GA, WA, SC, GU, MN, OK, NH, MD, ND, UT, HI, OR, LA, MS, KS, NM, ID, IA, VT and the countries of ARGENTINA, AUSTRIA, BELGIUM, BRASIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, KOREA, LATVIA, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH PACIFIC, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM.
- K numbers
- K212455
- Harmonized identifiers (openFDA)
- Device name
- Device, Neurovascular Embolization
- Medical specialty description
- Neurology
- Regulation number
- 882.5950
- Device class
- 2
- K number
- K951256, K000651, K143218, K073419, K012145, K140104, K112385, K200030, K073442, K042539, K203625, K242243, K093973, K052509, K063254, K112226, K020434, K083646, K222533, K242582, K041551, K962503, K914866, K233420, K172390, K050954, K160832, K023089, K002056, K071962, K060625, K094018, K960705, K920324, K103008, K120908, K033761, K161452, K173614, K901337 +147 more
- Registration number
- 3013353964, 9615728, 3015309643, 2134265, 3013944123, 1220477, 1225687, 3003524171, 1820334, 3008824097, 3009380063, 3008264254, 3009761573, 3008881809, 3000143502, 2011171, 3007628272, 9616684, 2029275, 3027595249, 3014644549, 3007738736, 3016591327, 3043869342, 3006950086, 2030598, 2032364, 3015615738, 3012536737, 3008853977, 3010173425, 3012931345, 3016245399, 2029214, 3008837339, 2022435, 9614654, 3015550451, 3005168196, 3022518360 +25 more
- FEI number
- 3013353964, 3015309643, 3013944123, 3000203466, 1220477, 3002913962, 3003524171, 1000518731, 1820334, 3008824097, 3009380063, 3008264254, 3009761573, 3003323921, 3008881809, 3000143502, 2011171, 3007628272, 3027595249, 3014644549, 3007738736, 3016591327, 3043869342, 3006950086, 3002534070, 3002947440, 3002808761, 3001451463, 3015615738, 1000306431, 3012536737, 3008853977, 3012931345, 3016245399, 3002095335, 3008837339, 2022435, 3015550451, 3005168196, 3022518360 +25 more