GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
GE Medical Systems SCS · GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client…
Open FDA's page for this record- Cfres ID
- 220684
- Product res number
- Z-2553-2026
- Event date initiated
- 2026-05-08
- Event date posted
- 2026-06-24
- Recall status
- Open, Classified
- Res event number
- 99042
- Product code
- LLZ
- Product description
- GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
- Code info
- REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384
- Firm FEI number
- 1000284313
- Recalling firm
- GE Medical Systems SCS
- Address 1
- 283 rue de la Miniere
- City
- Buc
- Country
- France
- Additional info contact
- GE HealthCare Service 800-437-1171
- Root cause description
- Software design
- Product quantity
- 340 servers
- Reason for recall
GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.
- Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 5/8/2026 by letter delivered via traceable means. The notice explained the issue, potential risk to patient, and requested the following: "Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use the device by following the instructions below: " Ensure that users verify the correct patient identity (including name, patient ID, and date of birth) and confirm that the selected study matches the intended patient and clinical context prior to reviewing or interpreting imaging studies. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via this FMI 80193 Digital CRF or print, fill out manually, scan, and email to Recall.80193@gehealthcare.com." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
- Distribution pattern
Worldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Morocco, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, United Arab Emirates, United Kingdom, Vietnam.
- K numbers
- K081985
- Harmonized identifiers (openFDA)
- Device name
- System, Image Processing, Radiological
- Medical specialty description
- Radiology
- Regulation number
- 892.2050
- Device class
- 2
- K number
- K103576, K103308, K032026, K020140, K201103, K050298, K063198, K130749, K141669, K073448, K092125, K163290, K113076, K090119, K111776, K140869, K162474, K171544, K141329, K253950, K202487, K081308, K896050, K162112, K223502, K121076, K170580, K211257, K121781, K040970, K043412, K954222, K173622, K143397, K220205, K212565, K233331, K914366, K032760, K082526 +2,257 more
- Registration number
- 3026612481, 3038191646, 3023266032, 3017233423, 3005207094, 8043933, 3002579136, 9613188, 3009719225, 3013427569, 3017585294, 1644408, 3030446518, 3022435840, 3019878658, 3009196021, 3009555440, 3025394753, 3025603301, 3013503781, 3017545364, 1054415, 9611343, 3017448360, 1045254, 1717344, 3007441485, 3009756153, 8022257, 3013682146, 3004555721, 3014232555, 3005156333, 3036725840, 3008487708, 3006165530, 3004722076, 9614587, 3005202320, 3030979155 +603 more
- FEI number
- 3026612481, 3038191646, 3023266032, 3017233423, 3005207094, 3002806725, 3002579136, 3003190524, 3009719225, 3013427569, 3017585294, 3030446518, 3022435840, 3019878658, 3009196021, 3007475226, 3009555440, 3025394753, 3025603301, 3013503781, 3017545364, 1000514697, 3017448360, 1045254, 1717344, 3007441485, 3009756153, 3013682146, 3004555721, 3014232555, 3005156333, 3036725840, 3008487708, 3006165530, 3004722076, 3005202320, 3030979155, 3017991646, 3010684280, 3019252669 +603 more