Device recalls

Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.

IPG Medical Corporation · Software anomaly that causes a false display of error code 5018.

Open, ClassifiedDevice DesignZ-2554-2026
Open FDA's page for this record
Cfres ID
220724
Product res number
Z-2554-2026
Event date initiated
2026-05-05
Event date posted
2026-06-24
Recall status
Open, Classified
Res event number
99064
Product code
GEX
Product description
Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.
Firm FEI number
3020221089
Recalling firm
IPG Medical Corporation
Address 1
225 Cedar Hill St
City
Marlborough
State
MA
Postal code
01752-5900
Additional info contact
Kevin O'Connell 774-633-5903
Reason for recall
Software anomaly that causes a false display of error code 5018.
Root cause description
Device Design
Product quantity
198 units
Distribution pattern
US Nationwide distribution in the state of MA.
Code info

Model No. MD2300F-001507; UDI-DI 00810071230226; Serial No. MDUF250342, MDUF250400, MDUF250429, MDUF250060, MDUF250061, MDUF250067, MDUF250068, MDUF250069, MDUF250071, MDUF250076, MDUF250083, MDUF250086, MDUF250091, MDUF250092, MDUF250104, MDUF250131, MDUF250133, MDUF250134, MDUF250159, MDUF250160, MDUF250199, MDUF250200, MDUF250202, MDUF250203, MDUF250204, MDUF250205, MDUF250240, MDUF250241, MDUF250242, MDUF250278, MDUF250297, MDUF250299, MDUF250312, MDUF250314, MDUF250325, MDUF250340, MDUF250341, MDUF250351, MDUF250352, MDUF250355, MDUF250367, MDUF250368, MDUF250370, MDUF250380, MDUF250384, MDUF250385, MDUF250395, MDUF250399, MDUF250409, MDUF250410, MDUF250427, MDUF250430, MDUF250431, MDUF250433, MDUF260009, MDUF260010, MDUF260011, MDUF260017, MDUF260019, MDUF260023, MDUF260024, MDUF260025, MDUF260026, MDUF260027, MDUF260032, MDUF260043, MDUF260067, MDUF260068, MDUF260070, MDUF260072, MDUF260077, MDUF260079, MDUF260094, MDUF260095, MDUF260096, MDUF260097, MDUF260098, MDUF260099, MDUF260100, MDUF260101, MDUF260103, MDUF260104, MDUF260105, MDUF260106, MDUF260107, MDUF260108, MDUF260109, MDUF260112, MDUF260115, MDUF250014, MDUF250015, MDUF250016, MDUF250050, MDUF250051, MDUF250052, MDUF250053, MDUF250054, MDUF250055, MDUF250056, MDUF250057, MDUF250058, MDUF250059, MDUF250062, MDUF250063, MDUF250064, MDUF250065, MDUF250066, MDUF250070, MDUF250072, MDUF250073, MDUF250074, MDUF250075, MDUF250084, MDUF250085, MDUF250088, MDUF250089, MDUF250090, MDUF250093, MDUF250102, MDUF250103, MDUF250105, MDUF250132, MDUF250145, MDUF250158, MDUF250161, MDUF250162, MDUF250198, MDUF250201, MDUF250206, MDUF250207, MDUF250220, MDUF250237, MDUF250238, MDUF250239, MDUF250243, MDUF250245, MDUF250279, MDUF250280, MDUF250295, MDUF250296, MDUF250298, MDUF250310, MDUF250311, MDUF250313, MDUF250323, MDUF250344, MDUF250348, MDUF250349, MDUF250350, MDUF250353, MDUF250357, MDUF250364, MDUF250365, MDUF250369, MDUF250371, MDUF250372, MDUF250378, MDUF250379, MDUF250381, MDUF250382, MDUF250383, MDUF250386, MDUF250393, MDUF250396, MDUF250397, MDUF250398, MDUF250407, MDUF250408, MDUF250411, MDUF250412, MDUF250428, MDUF250432, MDUF260016, MDUF260018, MDUF260020, MDUF260031, MDUF260041, MDUF260044, MDUF260069, MDUF260071, MDUF260078, MDUF260088, MDUF260089, MDUF260090, MDUF260091, MDUF260092, MDUF260093, MDUF250087, MDUF250208, MDUF250244, MDUF250324, MDUF250326, MDUF250354, MDUF250356, MDUF250394, MDUF260042, MDUF260045, MDUF260076; Software Version 2.2.

Action

An URGENT MEDICAL DEVICE RECALL notice dated 05/05/2026 was emailed to IPG Medical's consignee notifying them of this recall correction to address Error 5018 and Error 5007/5023 displayed on monitors with software version 2.2. The notification provides instructions to consignees to work around the error messages until a software upgrade is deployed at a later date. The notification instructs consignees that their staff may correct units onsite via software upgrade or units may be returned to IPG Medical for a factory correction when the update is available. A corrected IFU will be provided to all customers. Questions can be directed via phone to 774-633-5903 or koconnell@ipgmedical.com.

K numbers
K232568
Harmonized identifiers (openFDA)
Device name
Powered Laser Surgical Instrument
Medical specialty description
General, Plastic Surgery
Regulation number
878.4810
Device class
2
K number
K903154, K941838, K241358, K944472, K163449, K872179, K013941, K980517, K010284, K142093, K980838, K944346, K172376, K242251, K042139, K915352, K181614, K961279, K140964, K930361, K072925, K083082, K052991, K233024, K951117, K943479, K944206, K161162, K113140, K253828, K012398, K201466, K013366, K140470, K032667, K061898, K944629, K250261, K942729, K853901 +2,855 more
Registration number
3006985163, 3007731000, 3010027944, 3044175602, 2938071, 3017125161, 3001222017, 3003708273, 3009380063, 3006985208, 3012452798, 3010822033, 3042261306, 3016076698, 2134812, 3013121972, 2011068, 3007712511, 3011137372, 3034246188, 3009462521, 3011670002, 3017144492, 3004450661, 1032227, 3007085226, 3033715699, 3013404959, 3012294783, 1045254, 9611388, 3045892905, 3008776882, 3015392301, 3014310779, 1221543, 3015452693, 3044905918, 3012669557, 2032513 +453 more
FEI number
3006985163, 3007731000, 3010027944, 3044175602, 2938071, 3017125161, 3001222017, 3003708273, 3009380063, 3004143769, 3006985208, 3012452798, 3010822033, 3042261306, 3016076698, 3013121972, 2011068, 3007712511, 3000204219, 1000623228, 3008133910, 3034246188, 3007475226, 3009462521, 3011670002, 3017144492, 3004450661, 1032227, 3007085226, 3033715699, 3012294783, 1045254, 3045892905, 3008776882, 3015392301, 3014310779, 1221543, 3015452693, 3044905918, 3012669557 +451 more