Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.
IPG Medical Corporation · Software anomaly that causes a false display of error code 5018.
Open FDA's page for this record- Cfres ID
- 220724
- Product res number
- Z-2554-2026
- Event date initiated
- 2026-05-05
- Event date posted
- 2026-06-24
- Recall status
- Open, Classified
- Res event number
- 99064
- Product code
- GEX
- Product description
- Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.
- Firm FEI number
- 3020221089
- Recalling firm
- IPG Medical Corporation
- Address 1
- 225 Cedar Hill St
- City
- Marlborough
- State
- MA
- Postal code
- 01752-5900
- Additional info contact
- Kevin O'Connell 774-633-5903
- Reason for recall
- Software anomaly that causes a false display of error code 5018.
- Root cause description
- Device Design
- Product quantity
- 198 units
- Distribution pattern
- US Nationwide distribution in the state of MA.
- Code info
Model No. MD2300F-001507; UDI-DI 00810071230226; Serial No. MDUF250342, MDUF250400, MDUF250429, MDUF250060, MDUF250061, MDUF250067, MDUF250068, MDUF250069, MDUF250071, MDUF250076, MDUF250083, MDUF250086, MDUF250091, MDUF250092, MDUF250104, MDUF250131, MDUF250133, MDUF250134, MDUF250159, MDUF250160, MDUF250199, MDUF250200, MDUF250202, MDUF250203, MDUF250204, MDUF250205, MDUF250240, MDUF250241, MDUF250242, MDUF250278, MDUF250297, MDUF250299, MDUF250312, MDUF250314, MDUF250325, MDUF250340, MDUF250341, MDUF250351, MDUF250352, MDUF250355, MDUF250367, MDUF250368, MDUF250370, MDUF250380, MDUF250384, MDUF250385, MDUF250395, MDUF250399, MDUF250409, MDUF250410, MDUF250427, MDUF250430, MDUF250431, MDUF250433, MDUF260009, MDUF260010, MDUF260011, MDUF260017, MDUF260019, MDUF260023, MDUF260024, MDUF260025, MDUF260026, MDUF260027, MDUF260032, MDUF260043, MDUF260067, MDUF260068, MDUF260070, MDUF260072, MDUF260077, MDUF260079, MDUF260094, MDUF260095, MDUF260096, MDUF260097, MDUF260098, MDUF260099, MDUF260100, MDUF260101, MDUF260103, MDUF260104, MDUF260105, MDUF260106, MDUF260107, MDUF260108, MDUF260109, MDUF260112, MDUF260115, MDUF250014, MDUF250015, MDUF250016, MDUF250050, MDUF250051, MDUF250052, MDUF250053, MDUF250054, MDUF250055, MDUF250056, MDUF250057, MDUF250058, MDUF250059, MDUF250062, MDUF250063, MDUF250064, MDUF250065, MDUF250066, MDUF250070, MDUF250072, MDUF250073, MDUF250074, MDUF250075, MDUF250084, MDUF250085, MDUF250088, MDUF250089, MDUF250090, MDUF250093, MDUF250102, MDUF250103, MDUF250105, MDUF250132, MDUF250145, MDUF250158, MDUF250161, MDUF250162, MDUF250198, MDUF250201, MDUF250206, MDUF250207, MDUF250220, MDUF250237, MDUF250238, MDUF250239, MDUF250243, MDUF250245, MDUF250279, MDUF250280, MDUF250295, MDUF250296, MDUF250298, MDUF250310, MDUF250311, MDUF250313, MDUF250323, MDUF250344, MDUF250348, MDUF250349, MDUF250350, MDUF250353, MDUF250357, MDUF250364, MDUF250365, MDUF250369, MDUF250371, MDUF250372, MDUF250378, MDUF250379, MDUF250381, MDUF250382, MDUF250383, MDUF250386, MDUF250393, MDUF250396, MDUF250397, MDUF250398, MDUF250407, MDUF250408, MDUF250411, MDUF250412, MDUF250428, MDUF250432, MDUF260016, MDUF260018, MDUF260020, MDUF260031, MDUF260041, MDUF260044, MDUF260069, MDUF260071, MDUF260078, MDUF260088, MDUF260089, MDUF260090, MDUF260091, MDUF260092, MDUF260093, MDUF250087, MDUF250208, MDUF250244, MDUF250324, MDUF250326, MDUF250354, MDUF250356, MDUF250394, MDUF260042, MDUF260045, MDUF260076; Software Version 2.2.
- Action
An URGENT MEDICAL DEVICE RECALL notice dated 05/05/2026 was emailed to IPG Medical's consignee notifying them of this recall correction to address Error 5018 and Error 5007/5023 displayed on monitors with software version 2.2. The notification provides instructions to consignees to work around the error messages until a software upgrade is deployed at a later date. The notification instructs consignees that their staff may correct units onsite via software upgrade or units may be returned to IPG Medical for a factory correction when the update is available. A corrected IFU will be provided to all customers. Questions can be directed via phone to 774-633-5903 or koconnell@ipgmedical.com.
- K numbers
- K232568
- Harmonized identifiers (openFDA)
- Device name
- Powered Laser Surgical Instrument
- Medical specialty description
- General, Plastic Surgery
- Regulation number
- 878.4810
- Device class
- 2
- K number
- K903154, K941838, K241358, K944472, K163449, K872179, K013941, K980517, K010284, K142093, K980838, K944346, K172376, K242251, K042139, K915352, K181614, K961279, K140964, K930361, K072925, K083082, K052991, K233024, K951117, K943479, K944206, K161162, K113140, K253828, K012398, K201466, K013366, K140470, K032667, K061898, K944629, K250261, K942729, K853901 +2,855 more
- Registration number
- 3006985163, 3007731000, 3010027944, 3044175602, 2938071, 3017125161, 3001222017, 3003708273, 3009380063, 3006985208, 3012452798, 3010822033, 3042261306, 3016076698, 2134812, 3013121972, 2011068, 3007712511, 3011137372, 3034246188, 3009462521, 3011670002, 3017144492, 3004450661, 1032227, 3007085226, 3033715699, 3013404959, 3012294783, 1045254, 9611388, 3045892905, 3008776882, 3015392301, 3014310779, 1221543, 3015452693, 3044905918, 3012669557, 2032513 +453 more
- FEI number
- 3006985163, 3007731000, 3010027944, 3044175602, 2938071, 3017125161, 3001222017, 3003708273, 3009380063, 3004143769, 3006985208, 3012452798, 3010822033, 3042261306, 3016076698, 3013121972, 2011068, 3007712511, 3000204219, 1000623228, 3008133910, 3034246188, 3007475226, 3009462521, 3011670002, 3017144492, 3004450661, 1032227, 3007085226, 3033715699, 3012294783, 1045254, 3045892905, 3008776882, 3015392301, 3014310779, 1221543, 3015452693, 3044905918, 3012669557 +451 more