Device recalls

Surgify Halo, 4.0 mm, Extendable, Model/Catalog Number: 40.000.SEE.H1; drills, burrs, trephines & accessories (simple, powered)

SURGIFY MEDICAL OY · Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS).

Open, ClassifiedUnder Investigation by firmZ-2558-2026
Open FDA's page for this record
Cfres ID
220182
Product res number
Z-2558-2026
Event date initiated
2026-04-30
Event date posted
2026-06-24
Recall status
Open, Classified
Res event number
98918
Product code
HBE
Product description
Surgify Halo, 4.0 mm, Extendable, Model/Catalog Number: 40.000.SEE.H1; drills, burrs, trephines & accessories (simple, powered)
Code info
UDI: 06429811532076; All lots until the IFU update has been implemented
Firm FEI number
3015392173
Recalling firm
SURGIFY MEDICAL OY
Address 1
Otakaari 5 I 5 i otakaari
City
Espoo
Country
Finland
Reason for recall
Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS).
Root cause description
Under Investigation by firm
Product quantity
112 units
Distribution pattern
US Nationwide distribution in the states of MN, MA, IN, LA, CA, NY, FL.
Action

Surgify issued a Medical Device Correction Notice - Surgify Halo - Restriction of Use notice to its consignees on 4/30/2026 via email. The notice explained the issue, potential risk, and requested the following: "What steps must the user take? Please review your current practice and ensure the device is only used in accordance with the updated safety information. Instruction to Customers: The customers have been instructed to do the following: - Do not use this device in endoscopic surgical procedures. If endoscopic access is required, an alternative instrument should be selected. Specifically: . Do not use this device in endoscopic surgical procedures. If endoscopic access is required, an alternative instrument should be selected. . Ensure that all relevant clinical staff and surgical team members are made aware of this notice. . Please retain this notice for your records and attach it to the relevant product documentation. Please acknowledge receipt of this Medical Device Correction Notice by replying to raqa@surgifymedical.com within 30 business days, confirming that the notice has been read, understood, and communicated to all relevant users within your organization. Please include your facility name, your name, and your role. If you have previously used the device in endoscopic procedures without incident, no patient follow-up is required; however, this application should now be discontinued in line with the updated safety information." For questions regarding this matter, please contact us by phone at +358 105 176 310 or via email at support@surgifymedical.com.

K numbers
K251433
Harmonized identifiers (openFDA)
Device name
Drills, Burrs, Trephines & Accessories (Simple, Powered)
Medical specialty description
Neurology
Regulation number
882.4310
Device class
2
K number
K190172, K022942, K875224, K252060, K113476, K111520, K123270, K183203, K170442, K984035, K062842, K253627, K191479, K002903, K903664, K911812, K070876, K911819, K902703, K250380, K213072, K143320, K223199, K042783, K880601, K161704, K905097, K080802, K072315, K873692, K881418, K191049, K900388, K230619, K213454, K030172, K251433, K922299, K040076, K014060 +25 more
Registration number
3005706667, 2648623, 1649518, 3026776, 3009380063, 8043933, 8030607, 2027754, 3003120897, 1220724, 3015967359, 1226544, 3015440604, 3001239363, 3033509898, 8010468, 3025603301, 1319639, 3016172274, 1828288, 2650143, 3034676720, 3007997036, 3014725904, 1045254, 1836357, 1526534, 3010097171, 2183744, 3006128100, 9611253, 8010177, 3015151147, 2916714, 3006017180, 1833986, 2953359, 3014479313, 3009973505, 1625507 +101 more
FEI number
3005706667, 3008647423, 2648623, 3000203466, 3009380063, 3005601686, 3004169071, 2027754, 3003120897, 1220724, 3003064967, 3015967359, 3015440604, 3001239363, 3033509898, 3025603301, 3003755849, 1000185904, 3016172274, 1828288, 1000477302, 3034676720, 3007997036, 3014725904, 1045254, 3546, 1526534, 3010097171, 2183744, 3006128100, 3002619595, 3000205591, 3002807687, 3015151147, 2916714, 1000517638, 3006017180, 3014479313, 3009973505, 1625507 +101 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.