Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. DR G'S BREAST PACK DYNJ51124B LIPO SUPPLEMENT PACK-LF DYNJ0…

Medline Industries, LP · Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

Open, ClassifiedNo Marketing ApplicationZ-2584-2026
Open FDA's page for this record
Cfres ID
220727
Product res number
Z-2584-2026
Event date initiated
2026-02-27
Event date posted
2026-06-25
Recall status
Open, Classified
Res event number
99062
Product code
FTN
Firm FEI number
1417592
Recalling firm
Medline Industries, LP
Address 1
3 Lakes Dr
City
Northfield
State
IL
Postal code
60093-2753
Additional info contact
Haley Barclay 800-633-5463
Reason for recall
Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
Root cause description
No Marketing Application
Product quantity
113,843 kits
Product description

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. DR G'S BREAST PACK DYNJ51124B LIPO SUPPLEMENT PACK-LF DYNJ0341174D PLASTIC PACK DYNJ66746 DYNJ85702

Code info

DYNJ51124B UDI-DI 10193489579291 Lot 24ABS353 DYNJ0341174D UDI-DI 10195327437749 Lot 24BBB383 DYNJ66746 UDI-DI 10193489375527 Lot 24BBD114 DYNJ85702 UDI-DI 10195327536213 Lot 24ABP053

Action

This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.

Distribution pattern

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.

Harmonized identifiers (openFDA)
Device name
Plastic Surgery And Accessories Kit
Medical specialty description
General, Plastic Surgery
Regulation number
878.3925
Device class
1
Registration number
3016659819, 9611024, 3009158523, 3004111573, 1720734, 3007289408, 3015895045, 3003431869, 3009964461, 3045816907, 3005012805, 3045816902, 3004911384, 3006231901, 1030451, 3026197444, 3047515274, 3005595283, 1422634, 2431166, 9680357, 2029015, 3007279404, 3035642068, 3006389770, 2084346, 3039382823, 1423395, 3004519921, 3005032396, 3004122598, 1928237, 1047429, 2183650, 2433012, 3015173212
FEI number
3016659819, 3009158523, 3004111573, 1720734, 3007289408, 3015895045, 3003431869, 3009964461, 3045816907, 1000124817, 3005012805, 3002807170, 1000514656, 3045816902, 3004911384, 3006231901, 1030451, 3026197444, 3047515274, 3005595283, 3005846326, 1000160458, 3000717925, 3007279404, 3035642068, 3006389770, 3039382823, 2523865, 3005032396, 3004122598, 3004829691, 1928237, 1047429, 2183650, 2433012, 3015173212