Device recalls

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. KIT NEURO CSTM SAMP0551

Medline Industries, LP · Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

Open, ClassifiedNo Marketing ApplicationZ-2589-2026
Open FDA's page for this record
Cfres ID
220741
Product res number
Z-2589-2026
Event date initiated
2026-02-27
Event date posted
2026-06-25
Recall status
Open, Classified
Res event number
99062
Product code
FQM
Product description
Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. KIT NEURO CSTM SAMP0551
Code info
SAMP0551 UDI-DI 10195327268503 Lots 22JBG593 22JBJ361
Firm FEI number
1417592
Recalling firm
Medline Industries, LP
Address 1
3 Lakes Dr
City
Northfield
State
IL
Postal code
60093-2753
Additional info contact
Haley Barclay 800-633-5463
Reason for recall
Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
Root cause description
No Marketing Application
Product quantity
113,843 kits
Action

This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.

Distribution pattern

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.

Harmonized identifiers (openFDA)
Device name
Bandage, Elastic
Medical specialty description
General Hospital
Regulation number
880.5075
Device class
1
K number
K780231, K840107, K802992, K771758, K801990, K040936, K760139, K760138, K802015, K802789, K802785, K760142, K801458, K802784, K820141, K760140, K872868, K780570, K760141, K770958, K791211, K791256, K800280, K790210, K013244, K810137, K801532, K800197, K771563, K952738, K831551, K780606, K781931, K831728, K811173, K802991, K040054, K820094, K830622, K820142 +2 more
Registration number
3003887292, 9615662, 3008044484, 1519375, 2110420, 3006950086, 3013841116, 3030956456, 3011137372, 1721686, 3031757047, 3017220525, 3023809973, 3016872092, 3008520652, 3004014222, 3007048057, 3010043561, 9613778, 3005237621, 3015378469, 1721504, 3005034064, 3007282994, 3043226252, 3013986346, 3012236175, 3012865799, 3007966889, 3004488394, 3014146451, 1723702, 3016450032, 3016970619, 3019804885, 3031180770, 1649245, 3027481895, 2183164, 2320762 +665 more
FEI number
3000206450, 3003887292, 3008044484, 3002902735, 1519375, 2110420, 3006950086, 3013841116, 3030956456, 1721686, 3031757047, 3017220525, 3023809973, 3016872092, 3008520652, 3004014222, 3008439903, 3007048057, 3010043561, 3005237621, 3015378469, 1721504, 3005034064, 3007282994, 3043226252, 3013986346, 3012236175, 3012865799, 3007966889, 3004488394, 3014146451, 3016450032, 3016970619, 3019804885, 3031180770, 3027481895, 3004753358, 1720929, 3002858762, 3013919644 +665 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.