Device recalls

ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AI…

Tosoh Bioscience Inc · The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could c…

Open, ClassifiedComponent design/selectionZ-2746-2026
Open FDA's page for this record
Cfres ID
221069
Product res number
Z-2746-2026
Event date initiated
2026-06-12
Event date posted
2026-07-16
Recall status
Open, Classified
Res event number
99210
Product code
CDZ
Code info
UDI-DI: 14560189210032; Lot Numbers: FX12404, FY12407, FY12406, FZ12408
Firm FEI number
3005529799
Recalling firm
Tosoh Bioscience Inc
Address 1
3600 Gantz Rd
City
Grove City
State
OH
Postal code
43123-1895
Additional info contact
Isabelle Lang-Zwosta 615-8064671
Root cause description
Component design/selection
Product quantity
310 units
Distribution pattern
Global Distribution. US Nationwide.
Product description

ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204

Reason for recall

The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.

Action

Tosoh Bioscience Inc. sent an URGENT: MEDICAL DEVICE RECALL letter dated 6/9/26 to customers. Customers are to SEGREGRATE and quarantine all affected lots in inventory. Affected product lot should not be used. CONTACT Tosoh Customer Service department at 1-866-527-3587 to arrange return of the affected product. For questions, contact Tosoh Technical Support TESrecall@tosoh.com or 1-800-248-6764. For replenishment of returned Testosterone test cups, contact Customer Service department at 1-866-527-3587. Replenishments will be with ST AIA-PACK Testosterone Assay Test Cups from new lots. If you have obtained unexpected test results or received any complaints of illness or adverse events associated with the use of the ST AIA-PACK Testosterone Assay Test Cups and associated Testosterone reagents, contact Tosoh Technical Support at 1-800-248-6764 or TESrecall@tosoh.com Monday to Thursday 8:30AM-8:30PM EST Friday 8:30AM-7:30PM EST An URGENT: MEDICAL DEVICE RECALL: SECOND NOTIFICATION: dated 7/17/26 was sent to customers. The letter has additional language regarding the risks to health associated with the use of the affected lots. Instructions to customers are the same as in the first letter.

K numbers
K023091
Harmonized identifiers (openFDA)
Device name
Radioimmunoassay, Testosterones And Dihydrotestosterone
Medical specialty description
Clinical Chemistry
Regulation number
862.1680
Device class
1
K number
K791999, K955870, K811536, K964889, K802661, K091867, K841948, K802728, K915823, K772191, K800776, K201908, K883267, K915570, K955102, K830512, K955440, K813401, K001935, K021326, K133491, K120009, K014120, K152185, K854674, K240865, K802135, K233480, K001086, K252728, K924257, K191350, K102814, K840187, K191533, K941797, K854671, K122793, K904100, K803001 +49 more
Registration number
3002800697, 2032087, 9610126, 2432235, 3007361513, 8020888, 2032839, 1827821, 2245285, 8010132, 1219913, 3022178699, 9612316, 3005542422, 1832216, 2122870, 3007118747, 1319681, 1222302, 3005333358, 1036362, 3003559191, 8031673, 3005984081, 2032682, 1063851, 3002642396, 8023018, 9710337, 3009092698, 3003385757, 3007349751, 3004192065, 3006198300, 3002806944, 2029372, 8021924, 3005529799, 2018685, 1415939 +17 more
FEI number
3002806808, 3002800697, 2432235, 3007361513, 1827821, 2245285, 3002485711, 3003261083, 1219913, 3022178699, 2122870, 3005542422, 1832216, 1000442352, 3007118747, 3005333358, 1036362, 3003559191, 1000306617, 3005984081, 3000717645, 3003355251, 3007019572, 2967, 3002642396, 3004185580, 3009092698, 3003385757, 3007349751, 3004192065, 3006198300, 3002806944, 3002808140, 3002806559, 3005529799, 2018685, 1415939, 2419955, 3033967997, 2517506 +17 more