Brand Name: COAPTIUM Connect Product Name: COAPTIUM Connect Model/Catalog Number: NCOCO1510 Software Version: N/A Product Description: In s…
TISSIUM SA · Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use…
Open FDA's page for this record- Cfres ID
- 221370
- Product res number
- Z-2901-2026
- Event date initiated
- 2026-06-15
- Event date posted
- 2026-08-06
- Recall status
- Open, Classified
- Res event number
- 99345
- Product code
- SFD
- Code info
- Lot Code: Model No NCOCO1510; UDI-DI 03760279380070; Lot Numbers 250700041 and 251000032
- Firm FEI number
- 3021453307
- Recalling firm
- TISSIUM SA
- Address 1
- 74 Rue Du Faubourg Saint-Antoine
- Country
- France
- Additional info contact
- Hallie ARLOTTA 1-317-9957384
- Root cause description
- Incorrect or no expiration date
- Distribution pattern
- US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
- Product description
Brand Name: COAPTIUM Connect Product Name: COAPTIUM Connect Model/Catalog Number: NCOCO1510 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: N/A
- Reason for recall
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
- Action
On June 25, 2026, the firm began notifying consignees via direct visit by medical representatives to deliver the Customer Recall Notification Letter in person. Representatives removed affected units from customer shelves, collected completed response forms, and provided conforming replacement product on the same day.
- Harmonized identifiers (openFDA)
- Device name
- In Situ Polymerizing Peripheral Nerve Repair Device
- Medical specialty description
- Neurology
- Regulation number
- 882.5270
- Device class
- 2
- K number
- K251957, DEN240066
- Registration number
- 3010659146, 3010178296, 3002807295, 3010383847, 3021453307, 3015546200
- FEI number
- 3010659146, 3003110073, 3002807295, 3010383847, 3021453307, 3015546200