Device recalls

60 In Non-DEHP Minibore Extension Set RES ME2010 UDI-DI code: 10885403236235 MaxPlus Ext Set w/Needleless and Y-Site REF MPX5305-C UDI-DI c…

BD SWITZERLAND SARL · Due to mechanical issue with internal sterilization equipment resulting in potential underdosing of product during processing, which may have resulted in produ…

Open, ClassifiedProcess controlZ-2941-2026
Open FDA's page for this record
Cfres ID
221167
Product res number
Z-2941-2026
Event date initiated
2026-06-16
Event date posted
2026-08-13
Recall status
Open, Classified
Res event number
99264
Product code
FPA
Firm FEI number
3012798826
Recalling firm
BD SWITZERLAND SARL
Address 1
17 terre bonne park a4 route de crassier
City
Eysins
Country
Switzerland
Additional info contact
Adam Daniels 385-583-9999
Root cause description
Process control
Product quantity
93,700
Distribution pattern
U.S.: AZ, CA, CO, FL, GA, IL, IN, KS, MD, MI, MN, MO, MS, NC, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA O.U.S.: Canada
Product description

60 In Non-DEHP Minibore Extension Set RES ME2010 UDI-DI code: 10885403236235 MaxPlus Ext Set w/Needleless and Y-Site REF MPX5305-C UDI-DI code: 10885403274619 9 IN Extension Set w/Max Y REF MPX5300-C UDI-DI code: 10885403237218 The Intravascular Administration Set and Extension Set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the patient's artery or vein. The lntravascular Administration Set and Extension Set may incorporate componentry that aid in the prevention of accidental needle sticks.

Code info

60 In Non-DEHP Minibore Extension Set RES ME2010 UDI-DI code: 10885403236235 Lot Number: 26025260 MaxPlus Ext Set w/Needleless and Y-Site REF MPX5305-C UDI-DI code: 10885403274619 Lot Number: 26035455 9 IN Extension Set w/Max Y REF MPX5300-C UDI-DI code: 10885403237218 Lot Number: 26055528

Reason for recall

Due to mechanical issue with internal sterilization equipment resulting in potential underdosing of product during processing, which may have resulted in products not being sterilized correctly.

Action

On 06/16/2026, an "URGENT: Medical Device Recall" Letter was emailed to customers informing them that, BD has identified a mechanical issue with internal sterilization equipment resulting in potential underdosing of product during processing, which may have resulted in product not being sterilized correctly. Customer are instructed to: "Immediately discontinue use of the affected lot(s). "Check all inventory locations within your facility and destroy all affected units in accordance with your facility s process. If destruction is not immediately possible at your facility, then product should be quarantined until disposal. "If product is currently in use, replace with non-impacted product as soon as possible. 1. Provide a copy of this notification to any customers and/or facilities you may have distributed devices to. 2. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 3. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 4. Report any complaints experienced with the use of this product to BD via the North American Regional Complaint Center: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT Email: productcomplaints@bd.com For: Product Complaints, Technical Questions Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT

K numbers
K051499, K083472
Harmonized identifiers (openFDA)
Device name
Set, Administration, Intravascular
Medical specialty description
General Hospital
Regulation number
880.5440
Device class
2
K number
K931377, K781186, K020544, K123344, K830258, K221069, K180843, K892429, K884264, K212262, K891176, K242126, K864296, K153293, K994429, K781190, K041490, K791173, K843463, K760103, K771271, K941453, K030385, K964439, K170002, K882798, K080665, K984441, K992266, K852361, K060488, K001423, K171117, K964067, K990288, K974589, K954813, K143234, K013394, K023469 +1,135 more
Registration number
3003310007, 3018507299, 3003678543, 3005609751, 3013162291, 3047706961, 2029275, 3004111573, 3016096098, 3009882464, 3004122598, 1220973, 3004143450, 3017509841, 2032098, 3006891665, 1319211, 3010872792, 3003512059, 9610993, 8021079, 3036978618, 9610847, 3004519921, 9681572, 3009746425, 2024024, 3019004087, 3011642792, 3020417125, 1526611, 9614279, 3007007790, 1000523114, 1647137, 3008733163, 3012552122, 3003674698, 1061124, 3014246639 +391 more
FEI number
3003310007, 3018507299, 3003678543, 3005609751, 3013162291, 3047706961, 3003081151, 3004111573, 3016096098, 3009882464, 3004122598, 3001665800, 1220973, 1000518731, 1000112877, 3008002452, 1000160458, 3004143450, 3017509841, 3006891665, 3003886545, 1319211, 3002806588, 3001570557, 3010872792, 3003512059, 3005846326, 1000174591, 3036978618, 3000215273, 3009746425, 2024024, 3019004087, 3011642792, 3020417125, 1526611, 3007007790, 1000523114, 3008733163, 3012552122 +391 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.