Velys Enabling Tech Robotic-Assist Station REF: 451611001
Medos International Sarl · Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.
Open FDA's page for this record- Cfres ID
- 221615
- Product res number
- Z-3088-2026
- Event date initiated
- 2026-07-23
- Event date posted
- 2026-09-02
- Recall status
- Open, Classified
- Res event number
- 99467
- Product code
- OLO
- Product description
- Velys Enabling Tech Robotic-Assist Station REF: 451611001
- Firm FEI number
- 3008114965
- Recalling firm
- Medos International Sarl
- Address 1
- Chemin Blanc 36 36 chemin blanc
- City
- Le Locle
- Country
- Switzerland
- Additional info contact
- Shannon Rook +41 32 934 8472
- Reason for recall
- Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.
- Root cause description
- Under Investigation by firm
- Product quantity
- 29 stations
- Distribution pattern
- US Nationwide distribution in the states of DE, FL, IL, KY, MD, MI, MN, NJ, NM, NY, OR, PA, TN, TX, VA, WA.
- Code info
UDI: 10705034645995/Serial Numbers: W14316295301 W15316297501 W15316298001 W17316322901 W17316323301 W21316347401 W21316347501 W21316347601 W21316347801 W21316347901 W21316348001 W22316365801 W34316479001 W34316479101 W34316479201 W34316479301 W34316479401 W34316479601 W34316479801 W34316480001 W34316480101 W49316554301 W51316597001 W51316597101 Y03316603601 Y03316603701 Y03316603801 Y07316625901 Y15316690901
- Action
On July 23, 2026, DePuy Synthes contacted affected consigenees via phone call concerning the recall. On July 24, 2026 Depuy issued a Urgent Medical Device Correction notification via sales associate visit. DePuy ask consignees to take the following actions: 1. Do not use the subject VELYS ROBOTIC-ASSIST STATION and quarantine it immediately. The VELYS. 2. Navigation Station and VELYS Camera Station can continue to be used following the Instructions for Use. 3. Complete, sign, and return the attached Business Response Form to OneMD-Field-Actions@its.jnj.com within three (3) business days of receipt of this notification. 4. Forward this notice to anyone in your facility that needs to be informed 5. Post a copy of this notice in a visible area for awareness and keep a copy for your records. 6. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA's MedWatch Adverse Event Reporting Program Online: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reporting-program/reportingserious
- K numbers
- K233228
- Harmonized identifiers (openFDA)
- Device name
- Orthopedic Stereotaxic Instrument
- Medical specialty description
- Neurology
- Regulation number
- 882.4560
- Device class
- 2
- K number
- K183196, K252087, K203110, K121936, K121765, K182345, K200095, K182721, K212028, K122011, K250182, K172849, K201251, K193307, K193094, K083009, K210574, K243258, K202956, K251316, K261024, K221180, K230180, K162962, K202552, K263234, K181890, K093425, K143597, K140596, K213491, K220348, K230989, K233199, K250167, K191722, K162567, K252962, K172518, K193120 +404 more
- Registration number
- 3027153110, 9617297, 1226544, 2027487, 3009493875, 3010047402, 3007993775, 3007485169, 1314417, 3010009632, 3003139373, 3006846753, 3009554293, 3009871135, 3005819474, 3006563559, 3004727093, 3015869121, 3043543260, 3026576110, 2531477, 3036756245, 3008027912, 3034676720, 1218017, 3006654680, 3010039400, 1828288, 3015941638, 3014680149, 3005751028, 3001284227, 3024999441, 3010822102, 1421101, 3016443334, 3010120326, 3004893332, 3021902355, 2530808 +350 more
- FEI number
- 3027153110, 2027487, 3009493875, 3010047402, 3007993775, 1000139987, 3007485169, 1314417, 3007282753, 3010009632, 3003139373, 3002807231, 3006846753, 3009554293, 3009871135, 3005532635, 3005819474, 3006563559, 3003112432, 3004727093, 3008782585, 3015869121, 3043543260, 3026576110, 3002821504, 3002619595, 3036756245, 3008027912, 3034676720, 1218017, 3006654680, 3010039400, 1828288, 3005751028, 3014680149, 3001284227, 3024999441, 3010822102, 1421101, 3016443334 +350 more