Device recalls

Velys Enabling Tech Robotic-Assist Station REF: 451611001

Medos International Sarl · Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.

Open, ClassifiedUnder Investigation by firmZ-3088-2026
Open FDA's page for this record
Cfres ID
221615
Product res number
Z-3088-2026
Event date initiated
2026-07-23
Event date posted
2026-09-02
Recall status
Open, Classified
Res event number
99467
Product code
OLO
Product description
Velys Enabling Tech Robotic-Assist Station REF: 451611001
Firm FEI number
3008114965
Recalling firm
Medos International Sarl
Address 1
Chemin Blanc 36 36 chemin blanc
City
Le Locle
Country
Switzerland
Additional info contact
Shannon Rook +41 32 934 8472
Reason for recall
Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.
Root cause description
Under Investigation by firm
Product quantity
29 stations
Distribution pattern
US Nationwide distribution in the states of DE, FL, IL, KY, MD, MI, MN, NJ, NM, NY, OR, PA, TN, TX, VA, WA.
Code info

UDI: 10705034645995/Serial Numbers: W14316295301 W15316297501 W15316298001 W17316322901 W17316323301 W21316347401 W21316347501 W21316347601 W21316347801 W21316347901 W21316348001 W22316365801 W34316479001 W34316479101 W34316479201 W34316479301 W34316479401 W34316479601 W34316479801 W34316480001 W34316480101 W49316554301 W51316597001 W51316597101 Y03316603601 Y03316603701 Y03316603801 Y07316625901 Y15316690901

Action

On July 23, 2026, DePuy Synthes contacted affected consigenees via phone call concerning the recall. On July 24, 2026 Depuy issued a Urgent Medical Device Correction notification via sales associate visit. DePuy ask consignees to take the following actions: 1. Do not use the subject VELYS ROBOTIC-ASSIST STATION and quarantine it immediately. The VELYS. 2. Navigation Station and VELYS Camera Station can continue to be used following the Instructions for Use. 3. Complete, sign, and return the attached Business Response Form to OneMD-Field-Actions@its.jnj.com within three (3) business days of receipt of this notification. 4. Forward this notice to anyone in your facility that needs to be informed 5. Post a copy of this notice in a visible area for awareness and keep a copy for your records. 6. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA's MedWatch Adverse Event Reporting Program Online: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reporting-program/reportingserious

K numbers
K233228
Harmonized identifiers (openFDA)
Device name
Orthopedic Stereotaxic Instrument
Medical specialty description
Neurology
Regulation number
882.4560
Device class
2
K number
K183196, K252087, K203110, K121936, K121765, K182345, K200095, K182721, K212028, K122011, K250182, K172849, K201251, K193307, K193094, K083009, K210574, K243258, K202956, K251316, K261024, K221180, K230180, K162962, K202552, K263234, K181890, K093425, K143597, K140596, K213491, K220348, K230989, K233199, K250167, K191722, K162567, K252962, K172518, K193120 +404 more
Registration number
3027153110, 9617297, 1226544, 2027487, 3009493875, 3010047402, 3007993775, 3007485169, 1314417, 3010009632, 3003139373, 3006846753, 3009554293, 3009871135, 3005819474, 3006563559, 3004727093, 3015869121, 3043543260, 3026576110, 2531477, 3036756245, 3008027912, 3034676720, 1218017, 3006654680, 3010039400, 1828288, 3015941638, 3014680149, 3005751028, 3001284227, 3024999441, 3010822102, 1421101, 3016443334, 3010120326, 3004893332, 3021902355, 2530808 +350 more
FEI number
3027153110, 2027487, 3009493875, 3010047402, 3007993775, 1000139987, 3007485169, 1314417, 3007282753, 3010009632, 3003139373, 3002807231, 3006846753, 3009554293, 3009871135, 3005532635, 3005819474, 3006563559, 3003112432, 3004727093, 3008782585, 3015869121, 3043543260, 3026576110, 3002821504, 3002619595, 3036756245, 3008027912, 3034676720, 1218017, 3006654680, 3010039400, 1828288, 3005751028, 3014680149, 3001284227, 3024999441, 3010822102, 1421101, 3016443334 +350 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.