Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532…
Lutronic Corporation · Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in pati…
Open FDA's page for this record- Cfres ID
- 221941
- Product res number
- Z-3110-2026
- Event date initiated
- 2026-08-06
- Event date posted
- 2026-09-04
- Recall status
- Open, Classified
- Res event number
- 99601
- Product code
- GEX
- Firm FEI number
- 3004483538
- Recalling firm
- Lutronic Corporation
- Address 1
- 219 Sowon-Ro; Deogyang 219 Sowon-Ro Deogyang Goyang Korea the Republic of
- Root cause description
- Device Design
- Product quantity
- 1065
- Distribution pattern
- Worldwide distribution - US Nationwide and the countries of Canada and European Union.
- Product description
Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment
- Code info
Lot Code: Model: DermaV Software Version: 1.06 UDI #: 08800291820200 Serial Numbers: DV025A010, DV025A011, DV025A012, DV025A013, DV025A014, DV025A015 DV025A016, DV025A018, DV025A019, DV025A020, DV025A021, DV025A022, DV025B011, DV025B012, DV025B013, DV025B014, DV025B015, DV025D006, DV025D007, DV025D008, DV025D009, DV025D010, DV025E003, DV025E004, DV025E005, DV025E006, DV025E007, DV025E008, DV025E009, DV025E010, DV025E011, DV025E012, DV025E013, DV025E014, DV025E016, DV025E017, DV025F012, DV025F013, DV025F014, DV025F015, DV025F016, DV025G011, DV025G012, DV025G013, DV025G014, DV025G015, DV025G016, DV025G017, DV025G018, DV025G019, DV025G020, DV025J021, DV025J022, DV025J023, DV025J025, DV025J026, DV025J027, DV025J028, DV025J034, DV025J035, DV025J036, DV025J037, DV025J038, DV025J040, DV025K018, DV025K019, DV025K020, DV025K021, DV025K023, DV025K024, DV025K025, DV025K027, DV025M031, DV025M032, DV025M033, DV025M034, DV026B001, DV026B002, DV026B003, DV026B004, DV026B005, DV026B010, DV026B014, DV026B015, DV026B016, DV026B017, DV026B018, DV026C006, DV026C007, DV026C008, DV026C009, DV026C010, DV025M035
- Reason for recall
Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
- Action
Firm began notifying customers on 8/6/26. Customers were instructed to discontinue use of previous parameters and utilize the new revised clinical cryogen parameters. Updated cryogen parameters were provided with the letter. If you have questions or need assistance, please contact Clinical Support at 844-365-5060 or via email at clinical@cynosurelutronic.com.
- K numbers
- K203788
- Harmonized identifiers (openFDA)
- Device name
- Powered Laser Surgical Instrument
- Medical specialty description
- General, Plastic Surgery
- Regulation number
- 878.4810
- Device class
- 2
- K number
- K903154, K941838, K241358, K944472, K163449, K872179, K013941, K980517, K010284, K142093, K980838, K944346, K172376, K242251, K042139, K915352, K181614, K961279, K140964, K930361, K072925, K083082, K052991, K233024, K951117, K943479, K944206, K161162, K113140, K253828, K012398, K201466, K013366, K140470, K032667, K061898, K944629, K250261, K942729, K853901 +2,855 more
- Registration number
- 3006985163, 3007731000, 3010027944, 3044175602, 2938071, 3017125161, 3001222017, 3003708273, 3009380063, 3006985208, 3012452798, 3010822033, 3042261306, 3016076698, 2134812, 3013121972, 2011068, 3007712511, 3011137372, 3034246188, 3009462521, 3011670002, 3017144492, 3004450661, 1032227, 3007085226, 3033715699, 3013404959, 3012294783, 1045254, 9611388, 3045892905, 3008776882, 3015392301, 3014310779, 1221543, 3015452693, 3044905918, 3012669557, 2032513 +453 more
- FEI number
- 3006985163, 3007731000, 3010027944, 3044175602, 2938071, 3017125161, 3001222017, 3003708273, 3009380063, 3004143769, 3006985208, 3012452798, 3010822033, 3042261306, 3016076698, 3013121972, 2011068, 3007712511, 3000204219, 1000623228, 3008133910, 3034246188, 3007475226, 3009462521, 3011670002, 3017144492, 3004450661, 1032227, 3007085226, 3033715699, 3012294783, 1045254, 3045892905, 3008776882, 3015392301, 3014310779, 1221543, 3015452693, 3044905918, 3012669557 +451 more