Device recalls

Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Philips North America Llc · The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached.…

Open, ClassifiedRadiation Control for Health and Safety ActZ-3155-2026
Open FDA's page for this record
Cfres ID
222754
Product res number
Z-3155-2026
Event date initiated
2026-08-18
Event date posted
2026-09-29
Recall status
Open, Classified
Res event number
99865
Product code
JAK
Product description
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
Code info
UDI: (01)00884838101111
Firm FEI number
3016618143
Recalling firm
Philips North America Llc
Address 1
222 Jacobs St
City
Cambridge
State
MA
Postal code
02141-2289
Root cause description
Radiation Control for Health and Safety Act
Product quantity
62
Distribution pattern
U.S. and U.S. Territories.
Reason for recall

The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.

Action

Philips sent an Electronic Product Radiation Correction notification dated August 18, 2026 to the users of the affected systems to alert them of the issue and provide the actions that should be taken in order to prevent risks to patients. Philips will schedule an appointment with customers to install a software solution free of charge.

K numbers
K244008
Harmonized identifiers (openFDA)
Device name
System, X-Ray, Tomography, Computed
Medical specialty description
Radiology
Regulation number
892.1750
Device class
2
K number
K132222, K132087, K241200, K812400, K191777, K982060, K894887, K063184, K091195, K133640, K163341, K904971, K031871, K232206, K191550, K161157, K214019, K771990, K031206, K133589, K113548, K183046, K223311, K964583, K980169, K780115, K233749, K200524, K221727, K201231, K820688, K211180, K183291, K151383, K090991, K130645, K220961, K073153, K253450, K861708 +787 more
Registration number
3006894636, 3026391023, 3011416394, 3029521953, 1000246715, 2032112, 1317307, 3015060224, 8043933, 3027245002, 3004977335, 3009543051, 9612283, 3011802799, 1811755, 3001722928, 9613445, 3013875850, 3000126629, 3032109181, 3008355361, 9614698, 3004056159, 3003967489, 3018596740, 3010789868, 1034973, 3005930550, 9611343, 3015223550, 3018423337, 3014134519, 3032109225, 8022257, 3009154005, 3010281549, 2183744, 3005156333, 8043728, 3004091527 +125 more
FEI number
1000139922, 3006894636, 3026391023, 3011416394, 3029521953, 1000246715, 1317307, 3015060224, 3027245002, 3004977335, 3009543051, 1000284313, 3011802799, 1811755, 3001722928, 3013875850, 3000126629, 3032109181, 3008355361, 3004056159, 1000514697, 3003967489, 3018596740, 3010789868, 1034973, 3005930550, 3018423337, 3015223550, 3014134519, 3032109225, 3009154005, 3010281549, 2183744, 3002619595, 3005156333, 3004091527, 3006487372, 3031503890, 3011786147, 3014447948 +125 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.