VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

Biomerieux Inc · The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe,…

Open, ClassifiedPackaging process controlZ-3231-2026
Open FDA's page for this record
Cfres ID
222401
Product res number
Z-3231-2026
Event date initiated
2026-07-24
Event date posted
2026-09-18
Recall status
Open, Classified
Res event number
99749
Product code
LON
Product description
VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811
Code info
UDI-DI 03573026637712; Lot Number 1333393504
Firm FEI number
1950204
Recalling firm
Biomerieux Inc
Address 1
595 Anglum Rd
City
Hazelwood
State
MO
Postal code
63042-2320
Additional info contact
bioMerieux Customer Service Center 800-6822666
Root cause description
Packaging process control
Product quantity
97680 units
Distribution pattern
International distribution to the countries of Austria, Germany, Luxembourg.
Reason for recall

The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.

Action

An URGENT FIELD SAFETY NOTICE notification letter was sent to customers beginning 7/24/26. Required actions Dispose of any remaining inventory VITEK 2 AST-N461 TEST KIT (Ref 424811, Lot 1333393504) that were manufactured with Poucher 8 with timestamps from 10:06:00 to 11:11:59, as identified by product labeling. Inspect each pouch in inventory and dispose of cards with "P8" and timestamps between 10:06:00 to 11:11:59 on the label. Alternatively, dispose of all cards from Lot 1333393504 remaining in inventory. In accordance with your facility s policies, you may consider performing a retrospective review of clinically significant resistant results. Among tests previously performed, identify any possible false resistant results that may have occurred, analyze the related risks, and determine appropriate actions, if applicable. Please ensure that this information is communicated to all relevant personnel within your laboratory, retain a copy in your records, and forward this information to all parties that may use this product, including those to whom you may have transferred the product. Please complete the Acknowledgement Form in Attachment A and return it to your local bioMrieux representative to confirm receipt of this notice. It is important that you return the acknowledgement form to bioMrieux even if you determine that this product correction notice does not impact your facility. As a reminder, please visit the Resource Center to confirm if a reference package insert is associated with a Field Corrective Action or Field Safety Corrective Action. https://resourcecenter.biomerieux.com/ If you require additional assistance or have any questions, please contact your local bioMrieux Customer Service representative.

Harmonized identifiers (openFDA)
Device name
System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Medical specialty description
Microbiology
Regulation number
866.1645
Device class
2
K number
K093224, K210287, K031530, K024152, K040106, K050982, K042963, K121546, K163563, K183524, K070809, K050002, K063573, K040716, K023273, K190905, K043389, K032399, K080107, K060312, K073433, K132573, K051689, K163637, K023301, K023895, K193299, K033557, K023634, K053097, K020511, K063564, K063623, K220805, K063366, K093076, K111599, K050780, K191766, K230864 +251 more
PMA number
P820073, N50524, N50487, P800029, P870049, P810024, N50531, P840065, N50061, P800030, N50525, N50538
Registration number
3003030793, 3009171251, 3006198300, 3008007472, 2022807, 1318354, 3025772277, 2919016, 3013718871, 3025775667, 1423662, 1950204, 1111096, 1119779, 3002444944, 3009637528, 3009288740, 1833407, 3007502289, 3027320620, 3031977134, 3007344102
FEI number
3003030793, 3006198300, 3008007472, 2022807, 1318354, 3006759904, 3025772277, 2919016, 3013718871, 3025775667, 3000215273, 1950204, 1111096, 1119779, 3002444944, 3009637528, 1000518597, 3009288740, 3007502289, 3027320620, 3031977134, 3007344102