EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with soft…

Bard Peripheral Vascular Inc · Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pre…

Open, ClassifiedSoftware designZ-3275-2026
Open FDA's page for this record
Cfres ID
221919
Product res number
Z-3275-2026
Event date initiated
2026-08-20
Event date posted
2026-09-25
Recall status
Open, Classified
Res event number
99590
Product code
KNW
Product description
EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with software
Firm FEI number
2020394
Recalling firm
Bard Peripheral Vascular Inc
Address 1
850 W Rio Salado Pkwy
City
Tempe
State
AZ
Postal code
85281-3802
Additional info contact
844-823-5433
Root cause description
Software design
Product quantity
99
Distribution pattern
US Nationwide distribution in the states of KS, SC, CA, TX, GA, NC, PA, SC, ID, MI, KY, CT, IL, VA, MA, WI, NY, IN, TN, OH, UT.
Code info

Software Versions 1.1.1.10207 and 1.2.0.12133. UDI-DI: 00801741235221 and 00801741234248. Serial Numbers: VTLNPX06, VTLNPX0F, VTLNPX0D, VTLNPX08, VTLNPX0K, VTLNPX0A, VTLNPX0C, VTLNPX0L, VTLNPX0J, VTLNPX04, VTLNPX0H, VTLNPX02, VTLPPX1G, VTLPPX1D, VTLPPX1C, VTLPPX1M, VTLPPX2H, VTLPPX2I, VTLPPX2F, VTLQPX15, VTLQPX1Q, VTLQPX1G, VTLQPX14, VTLQPX0T, VTLQPX1Y, VTLQPX1M, VTLQPX2G, VTLQPX1O, VTLQPX1K, VTLQPX19, VTLQPX2E, VTLQPX1W, VTLQPX10, VTLQPX1Z, VTLQPX24, VTLQPX21, VTLQPX0X, VTLQPX1E, VTLQPX0U, VTLQPX26, VTLQPX2F, VTLQPX25, VTLQPX0Z, VTLQPX18, VTLQPX28, VTLQPX20, VTLQPX27, VTLQPX11, VTLQPX2B, VTLQPX1F, VTLQPX2H, VTLQPX41, VTLQPX3T, VTLQPX43, VTLRPX0R, VTKYPX09, VTKYPX44, VTKYPX46, VTKYPX47, VTKYPX48, VTKYPX49, VTKZPX14, VTKZPX15, VTKZPX20, VTKZPX17, VTLNPX4I, VTLNPX49, VTLNPX4G, VTLNPX4N, VTLNPX4E, VTLNPX4L, VTLNPX44, VTLNPX46, VTLNPX4C, VTLPPX27, VTLPPX28, VTLPPX29, VTLPPX1Z, VTLPPX37, VTLPPX3B, VTLPPX34, VTLPPX38, VTLPPX36, VTLQPX0Q, VTLQPX0B, VTLQPX0E, VTLQPX0P, VTLQPX0F, VTLRPX0X, VTLRPX0Z, VTKYPX07, VTKYPX06, VTKYPX04, VTLNPX4B, VTLNPX42, VTLNPX4F, VTLPPX3D, VTLQPX0M, VTKYPX51

Reason for recall

Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pressure sensor (code E01-0054000E), which prompts the user to restart and re-calibrate; errors may lead to interruption or prolongation of the biopsy procedure, or the need for additional tissue sampling or repeat biopsy.

Action

On 8/20/2026, correction notices were mailed to customer who were asked to do the following: 1.If calibration cannot be completed, clinicians should first follow the troubleshooting instructions provided on the screen and in the Instructions for Use (IFU). If the issue persists, the probe should be replaced and the process repeated. If unsuccessful, an alternative biopsy driver or system should be used, as clinically appropriate. 2. If a pressure sensor error is observed, clinicians should follow the troubleshooting instructions provided on the screen and in the Instructions for Use (IFU). If the error occurs and the system cannot be returned to normal operations, discontinue use of the affected system and complete or reschedule using alternative equipment, as clinically appropriate. No additional follow-up is recommended for patients whose procedures were completed without complication. 3. If affected devices function as intended or if the error codes can be cleared, facilities can continue to use them prior to software update. 4. Post the correction notice on or near affected devices until the software has been updated. 5. Ensure devices are available for the update. 6. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Share this notification with all product users within your facility network to ensure they are also aware of this product correction. If the affected products have been distributed further, please ensure that these entities are contacted and informed of the correction. 7. Complete and return the response form via email to BDRC33@bd.com Firm has developed a software solution to address these issues. A firm representative will contact your facility to arrange the update. If you require assistance, contact the firm via email: productcomplaints@bd.com, or phone: 1-844-823-5433

K numbers
K252681
Harmonized identifiers (openFDA)
Device name
Instrument, Biopsy
Medical specialty description
Gastroenterology, Urology
Regulation number
876.1075
Device class
2
K number
K962977, K112411, K111100, K011206, K974158, K980248, K093512, K904658, K894419, K141910, K962907, K930083, K180807, K092338, K142477, K060413, K012088, K935814, K050795, K063233, K914874, K940365, K851837, K974575, K211933, K950732, K961407, K190378, K960064, K012023, K160316, K883374, K192948, K060384, K080171, K020987, K963825, K933448, K010400, K032847 +325 more
Registration number
3014032451, 3030446844, 3018094310, 3042940295, 3008044552, 3004122598, 3006788678, 3007648417, 1226544, 9617616, 3021023752, 1835568, 3007748770, 1000484830, 3002806603, 3019807891, 2531475, 3030626857, 3022977243, 9617592, 3007162296, 3004962788, 3007137598, 1061124, 3006981798, 3011504993, 3044399745, 3007664053, 1221485, 2134812, 1722746, 3009050824, 3015980719, 3004111573, 3011245468, 3015806723, 3011137372, 3033589330, 2953359, 3044026004 +166 more
FEI number
3014032451, 3030446844, 3018094310, 3042940295, 3008044552, 3004122598, 3006788678, 3007648417, 3021023752, 3007748770, 1000484830, 3002806603, 3019807891, 3030626857, 3002807847, 3022977243, 3007162296, 3004962788, 3007137598, 3006981798, 3011504993, 3044399745, 3007664053, 1221485, 1000434163, 3003015552, 3009050824, 1000123143, 3015980719, 3004111573, 3011245468, 3015806723, 3033589330, 3000206435, 3044026004, 3001236905, 2246552, 3006786907, 3000209794, 3015924373 +166 more
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.