Computed Tomography X-ray System, Model: uCT 530 with software
Shanghai United Imaging Healthcare Co., Ltd. · Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are…
Open FDA's page for this record- Cfres ID
- 222317
- Product res number
- Z-3277-2026
- Event date initiated
- 2026-08-11
- Event date posted
- 2026-09-25
- Recall status
- Open, Classified
- Res event number
- 99731
- Product code
- JAK
- Product description
- Computed Tomography X-ray System, Model: uCT 530 with software
- Firm FEI number
- 3011015597
- Recalling firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Address 1
- No. 2258 Chengbei Rd, Jiading Ind. Chengbei Rd
- Address 2
- No. 2258
- City
- Shanghai
- Country
- China
- Root cause description
- Software design
- Product quantity
- 4
- Code info
UDI-DI: 06971576831012. Serial Numbers/Software: 300174/R001.19.5.1839204-Re3300SP1H1_P4, 304006/R001.17.6.1607336-Re3100SP4, 304007/R001.19.4.1810621-Re3300SP1H1_P3, 304008/R001.19.4.1810621-Re3300SP1H1_P3
- Reason for recall
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
- Action
On 8/11/2026, correction notices were emailed to customers informing them that software will be released to resolve the software problem. Customers were asked to do the following: When using the Study Split function, follow the requirements below: 1) Temporarily disable the "Study Split" function. Navigate to the system configuration interface and uncheck the "Auto Study Split" option. 2) If the "Study Split" function is clinically necessary, open the system configuration interface and select "Split on PACS" instead of "Split on Local". 3) When "Auto study split" is checked, ensure that PACS (Picture Archiving and Communication System) is configured on site and auto-archive is enabled. 4) If images are found to be missing on the PA interface, immediately verify whether the corresponding image data exists in the PACS. a) If images exist in PACS: They can be retrieved from PACS for review. b) If images do not exist in PACS: Contact the firm's service engineer for assistance. For further questions, contact firm service engineer or call firm Customer Contact Center at (855) 221-1552 or 832-699-2799
- Distribution pattern
US Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX, ID, LA, MT, FL, CO, IN, KY, AL, OH, MI, IA, MO, NY, NV, TN, WV, SC, WI, IL, MS, MA, NJ.
- K numbers
- K181414, K200016
- Harmonized identifiers (openFDA)
- Device name
- System, X-Ray, Tomography, Computed
- Medical specialty description
- Radiology
- Regulation number
- 892.1750
- Device class
- 2
- K number
- K061310, K090462, K884668, K213999, K120833, K972903, K133640, K970980, K103489, K973168, K201231, K131431, K964583, K082458, K073373, K003408, K152584, K093891, K171261, K091529, K151414, K760688, K011160, K882378, K964890, K771978, K080271, K161285, K781431, K910859, K081482, K150713, K913485, K160743, K211591, K961055, K030450, K070658, K150785, K191891 +786 more
- Registration number
- 3004962788, 3015763655, 3016248538, 3003202425, 3008632256, 3027245002, 8043933, 3016965352, 3006487372, 3010398867, 3005810333, 3017426576, 2030624, 3012957335, 3033527282, 3049422862, 3026391023, 2134243, 3002808157, 3009077524, 3008284470, 3027373189, 3018094310, 3012761544, 3008007310, 9612283, 2126677, 3004198450, 3007797756, 3019216, 1226188, 3017375271, 3002907620, 1000246715, 9613299, 3027339477, 3003575350, 9617978, 3005156333, 3008769756 +127 more
- FEI number
- 1000139922, 3004962788, 3015763655, 3016248538, 3002619595, 3003007876, 3003202425, 3008632256, 3027245002, 3003076831, 3006487372, 3010398867, 3005810333, 3017426576, 3016965352, 3003079232, 3012957335, 3033527282, 3049422862, 3026391023, 3002808157, 3009077524, 3008284470, 3027373189, 3018094310, 3012761544, 3008007310, 2126677, 3004198450, 3007797756, 3002807903, 3019216, 3017375271, 3002907620, 1000514697, 1000246715, 3027339477, 3003575350, 3005156333, 3008769756 +127 more