Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/Catalog Number: HX55W-S0 Software V…

SPECTRUM MEDICAL SRL · Due to blood-to-glycol leak in the heat exchanger.

Open, ClassifiedProcess designZ-3328-2026
Open FDA's page for this record
Cfres ID
221606
Product res number
Z-3328-2026
Event date initiated
2026-07-17
Event date posted
2026-09-29
Recall status
Open, Classified
Res event number
99462
Product code
DTR
Code info
Lot Code: 18051160301178 (HX55W-S0)
Firm FEI number
3016746283
Recalling firm
SPECTRUM MEDICAL SRL
Address 1
Via Di Mezzo 23 23 via di mezzo
City
Mirandola
Country
Italy
Additional info contact
Cameron Schmitt 1-803-8333964
Reason for recall
Due to blood-to-glycol leak in the heat exchanger.
Root cause description
Process design
Product quantity
3232
Distribution pattern
US Nationwide distribution in the states of AL, CA, FL, KS MA, MD, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WI.
Product description

Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/Catalog Number: HX55W-S0 Software Version: N/A Product Description: The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 8 l/min. Quantum PureFlow Standard Heat Exchanger is a single use device which allows heating/cooling the blood during extracorporeal circulation. The device is sold sterile in a blister pack in a multiple box of 4 units. Device surfaces in contact with blood are treated with a phosphorylcholine-based biocompatible coating. The device has been designed to be powered by heaters coolers that use water or a glycol-based heat transfer fluid (HTF). Component: N/A

Action

On 07/17/2026, an "URGENT MEDICAL DEVICE RECALL" letter was sent via email to customers informing them that, Spectrum Medical S.r.l. has received customer complaints reporting a leak between the blood and glycol compartments of certain adult heat exchangers. Actions to be taken by customers: During the procedure, the user should monitor the heat exchanger for leakage, which would be evident by a change in colour in the glycol compartment. Below are images of the heat exchanger if a blood-to-glycol leak was to occur. When operating as intended, the glycol compartment should remain clear. If the colour of the glycol compartment changes, then a leak is present between the blood and glycol compartments. If leakage is noted between the blood and glycol compartments of the heat exchanger, the heat exchanger should be immediately replaced in the circuit with a new device. Please return the defective heat exchanger to Spectrum Medical. The contaminated heater-cooler device should also be immediately removed from clinical use and returned to Spectrum Medical for decontamination and inspection. Do not use the heater-cooler with another patient. If leakage has occurred, please contact a Spectrum Medical representative to coordinate device return. If leakage into the water compartment occurs when using a third-party heater-cooler, the customer should immediately remove the heater-cooler from clinical use and contact the heater-cooler manufacturer. Spectrum Medical suggests that for customers that use the heat exchanger in combination with a third-party heater-cooler to contact the manufacturer of the heater-cooler to pre-emptively discuss the steps that should be taken if the device is exposed to blood contamination. For questions or further assistance - contact Regulatory Affairs Specialist at cameron.schmitt@spectrummedical.com

K numbers
K220110, K223879
Harmonized identifiers (openFDA)
Device name
Heat-Exchanger, Cardiopulmonary Bypass
Medical specialty description
Cardiovascular
Regulation number
870.4240
Device class
2
K number
K000977, K964571, K953838, K884560, K973475, K944267, K823451, K904171, K883331, K840256, K853410, K212688, K896807, K905500, K781210, K781764, K220110, K883439, K913691, K802879, K873561, K944219, K040355, K201984, K931967, K240908, K900640, K934355, K914742, K770017, K910923, K800377, K020106, K223879, K813264, K853516, K870804, K811204, K770018, K252541 +51 more
Registration number
2916596, 1724474, 3010155661, 3016746283, 1124841, 3019807891, 3002806603, 3003752502, 3017987980, 3030125051, 3002807090, 3015136750, 1218444, 3011554160, 3010163695, 1625519, 3013267343, 1319639, 3006073153, 2184009, 8010762, 3044156918, 9681834, 1928237, 9680841, 8010026, 1643817, 2029275, 1649914, 3005941719, 9617601, 2011171, 1721676
FEI number
2916596, 3002808327, 3010155661, 3009649876, 2828, 3016746283, 3003094851, 3019807891, 3002806603, 3003752502, 3000204839, 3017987980, 1000306431, 3030125051, 3015136750, 1218444, 3011554160, 3005426550, 3010163695, 3002807111, 1625519, 3013267343, 1000116158, 3006073153, 3006003617, 3001665800, 3044156918, 1928237, 1643817, 1000138054, 3002534070, 1721676, 2011171
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.