Device recalls

Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 S…

Tandem Diabetes Care, Inc. · A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application w…

Open, ClassifiedSoftware Design ChangeZ-3330-2026
Open FDA's page for this record
Cfres ID
221918
Product res number
Z-3330-2026
Event date initiated
2026-07-16
Event date posted
2026-09-29
Recall status
Open, Classified
Res event number
99589
Product code
QFG
Code info
Lot Code: UDI-DI: 00389152470641 Software Version: 2.13 (iOS only)
Firm FEI number
3007981285
Recalling firm
Tandem Diabetes Care, Inc.
Address 1
12400 High Bluff Dr
City
San Diego
State
CA
Postal code
92130-3077
Additional info contact
Susan Morrison 1-858-3666880
Root cause description
Software Design Change
Product quantity
177
Distribution pattern
US Nationwide distribution.
Product description

Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No

Reason for recall

A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.

Action

On 07/16/2026, an "URGENT MEDICAL DEVICE CORRECTION" communication was sent via certified mail to customers informing them that an unexpected behavior was identified in the t:slim iOS mobile application version 2.13 made available through TestFlight. The issue involves users unable to pair a compatible FreeStyle Libre 3 Plus sensor after updating to version 2.13. If this occurs, the application will incorrectly display an "Incompatible Sensor" message and prevent sensor use. Customer are advised to: 1. Discontinue use of iOS t:slim mobile application version 2.13 2. Uninstall version 2.13 from their mobile device. 3. Return to the current production version of the t:slim mobile application (version 2.12.1) available on the Apple App Store. 4. Scan the compatible FreeStyle Libre 3 Plus sensor with version 2.12.1 when starting their next sensor session. For Questions - contact Tandum Customer Support at 1-877-801-6901 or email Techsupport@tandumdiabetes.com (24/7/365).

K numbers
K201214, K203234, K232380
Harmonized identifiers (openFDA)
Device name
Alternate Controller Enabled Insulin Infusion Pump
Medical specialty description
Clinical Chemistry
Regulation number
880.5730
Device class
2
K number
K213134, K201214, K261672, K203234, K232380, K213536, K233044, K241078, K231485, K253074, K223846, K191679, K250930, K231826, K250792, K230545, K253743, K241178, K253470, K233952, K240309, K234027, K252770, K260001, K253976, K203768, K263422, K223213, K251032, DEN180058
Registration number
3014197758, 3010291427, 2030624, 3032746481, 3030970576, 1220973, 3019807891, 2024024, 3013756811, 3016798778, 3016540876, 3008824097, 3026266833, 3014585508, 2032227, 2032098, 3044561902, 2030598, 1220477, 3015192982, 2246552, 2022435, 3016245399, 3019004087, 3007981285, 2029275, 3047799708, 3005627859, 3042027705
FEI number
3014197758, 3010291427, 3032746481, 3030970576, 1220973, 3019807891, 2024024, 3013756811, 3016798778, 3016540876, 3008824097, 3026266833, 3014585508, 1000306431, 3044561902, 3003058448, 3000206435, 1220477, 3015192982, 1000518731, 2246552, 3003166194, 2022435, 3016245399, 3019004087, 3007981285, 3047799708, 3005627859, 3042027705
Wholesale access for licensed clinicians.

Free to join: your NPI is checked against the federal registry at signup, then wholesale pricing and the professional shelves open.

Data from openFDA, FDA's public API. FDA says not to rely on it to make decisions about medical care and to treat every result as unvalidated; confirm against FDA's own record and the product labeling. Not medical or legal advice.